Question 1

During a Holter hookup at a multispecialty clinic, a 67-year-old woman says she is flying to visit her sister tomorrow and asks what she should stay away from while she wears the recorder for 48 hours. Which list should the technician give her?

  1. Sleeping on her side, carrying a purse and wearing a bra
  2. Caffeine, nicotine, alcohol and over-the-counter cold remedies
  3. Stair climbing, brisk walking, lifting and household chores
  4. Metal detectors, magnets, electric blankets and high-voltage areas
Show the answer and rationale

Answer: D

  • A. Keeping the recorder close to the body may make sleep less comfortable, but positions and ordinary clothing do not harm the recording when the electrodes stay attached.
  • B. These can affect heart rhythm, but they are not device precautions, and advice about drugs or habits comes from the ordering provider, not the technician.
  • C. Patients are asked to keep up their usual activities, because the test aims to record a typical day, including the exertion that may bring on symptoms.
  • D. Correct. These can interfere with the recording or the device, so she should avoid them for the length of the test.

Strong magnetic and electrical fields can add noise or disrupt an ambulatory recorder, so patients wearing a Holter avoid magnets, metal detectors, electric blankets and high-voltage areas, and keep the device dry. Otherwise they continue normal activity and keep a timed diary. If tape or adhesive allergy exists, they say so before the hookup, and any skin reaction afterward is reported to the office that placed the monitor.

Sources: LITFL ECG Library

Question 2

In a Holter scanning lab, a technician summarizes a 24-hour recording from a 66-year-old patient with heart failure and a biventricular (CRT) pacemaker. Data are: 100,000 total beats; 16,000 PVCs; each remaining beat biventricular paced with capture; no pauses; no runs of VT. The device withholds pacing on each PVC and resumes on the next cycle. Which statement BEST summarizes the device-related finding for the device clinic?

  1. Oversensing that inhibits pacing at each PVC
  2. Intermittent failure to capture during PVCs
  3. Biventricular pacing below target at 84%
  4. Normal device function with incidental PVCs
Show the answer and rationale

Answer: C

  • A. Oversensing is the device reacting to signals that are not heartbeats. A PVC is a real ventricular beat, so sensing it and withholding output is correct sensing.
  • B. Failure to capture means a spike with no QRS. A PVC is the patient's own beat, and the device correctly holds its output, so no spike is missing its capture.
  • C. Correct. 100,000 − 16,000 = 84,000 paced beats, or 84%, well under the >95% target; each PVC replaces a resynchronized beat, so the finding is reported to the device clinic.
  • D. The device is sensing and pacing correctly, which makes this tempting, but in a CRT patient frequent PVCs matter because every PVC is a beat without biventricular pacing.

Cardiac resynchronization therapy helps only when the device paces both ventricles on nearly every beat. The 2023 HRS/EHRA remote device clinic consensus states that response tracks the percentage of biventricular pacing and that more than 95% should be the target, with alerts so lost pacing can be restored quickly; AF is a common cause. The app's device sheet adds that PVCs or AF that cut the percentage are reported. Here 16,000 of 100,000 beats were PVCs, so only 84% were paced. The device is not malfunctioning; the PVC burden is the finding the device clinic needs.

Sources: HRS/EHRA/APHRS/LAHRS Remote Device Clinic Consensus (2023); CanadiEM: Interpreting a Pacemaker ECG (2016)

Question 3

In a Holter clinic, a technician hands a patient who is blind the printed instructions and symptom diary for a 24-hour recording. The patient says, 'I can't read these.' Which action is BEST?

  1. Read both aloud in full, and ask the nurse about an accessible way to keep the diary
  2. Send the sheets home so a family member can read them to her later
  3. Summarize the main points aloud and skip the diary, since she cannot read the form
  4. Read the key steps aloud, then have her sign that she understood
Show the answer and rationale

Answer: A

  • A. Correct. The patient gets the complete information, and the nurse can arrange another way to keep the diary.
  • B. This delays the teaching until after the monitor is on, and a relative may not be available or may misread a step.
  • C. Tempting because it saves time. A summary leaves out content, and dropping the diary removes the symptom log that makes the recording useful.
  • D. A signature records that she was told, not that she can follow every step, and the key steps leave out parts of the instructions and the diary.

The ADA's effective communication rules call for accessible ways to share printed information with people who are blind, such as a qualified reader, Braille, large print or electronic formats, and the reader must read effectively, accurately and impartially. Here the technician reads the content in full, then involves the nurse to arrange an accessible way to log symptoms. Teach-back, asking the patient to say the steps in her own words, confirms she can follow them while wearing the monitor.

Sources: ADA.gov: ADA Requirements, Effective Communication

Question 4

At an internal medicine clinic, a 61-year-old calls 30 hours into a 48-hour Holter recording. He says the recorder turned off this morning and will not turn back on, although the electrodes are still on his chest. Which instruction should the technician give him?

  1. Keep wearing it and return it at 48 hours as planned
  2. Remove it now, since 30 hours is enough
  3. Bring the recorder in today so it can be checked
  4. Write any symptoms in the diary until 48 hours end
Show the answer and rationale

Answer: C

  • A. Sticking to the schedule sounds sensible, but a recorder that is off captures nothing, so the extra hours of wear add no data.
  • B. The order was for 48 hours, and whether a shorter recording is enough is the provider's decision, not the technician's.
  • C. Correct. A recorder that has turned off before the test ends is no longer recording, so the clinic needs to check it and let the provider decide how to complete the test.
  • D. Diary entries matter only when there is a recording to match them with. Without one, the reader cannot tell which rhythm went with each symptom.

A Holter monitor records the ECG continuously for 24 to 48 hours during normal activity, and patients are told to contact the office if the monitor turns off before the test ends. The technician has the patient bring the recorder in so the battery, memory and connections can be checked, notes the time it stopped, and tells the ordering provider. The provider then decides whether the hours already recorded are enough or whether the test should be restarted or repeated.

Sources: LITFL ECG Library

Question 5

In a cardiology office, a 63-year-old with spells of fluttering in his chest is sent home with a handheld symptom event recorder. He will carry it in a pocket, and it has no wires that stay attached to his skin. Which instruction should the technician give for when a spell begins?

  1. Call the monitoring center so staff can start it
  2. Press its button after the spell has ended
  3. Hold it on your chest and start recording
  4. Write the time down and record the next spell
Show the answer and rationale

Answer: C

  • A. Staff cannot switch on a device in his pocket. Calling first delays the recording, and brief spells may end before it starts.
  • B. This is the instruction for a looping monitor that is worn continuously and saves the ECG from just before the button press. A carried symptom recorder has no stored ECG to save.
  • C. Correct. A symptom event recorder captures nothing until it is placed against the chest and turned on, so the patient must do this while the fluttering is happening.
  • D. Diary notes help, but waiting for a later spell misses the rhythm during this one. The recording should be made while symptoms are present.

Event monitors differ in how they capture a spell. A loop memory monitor stays attached by electrodes, records continuously and saves the stretch before, during and after a button press. A symptom event recorder is carried, then placed on the chest and switched on only when symptoms occur, so it records only what happens after it is applied. Because of that, it works for spells that last long enough to apply the device, and the technician should teach the patient to act quickly, keep a diary and follow the steps for sending the recording.

Sources: LITFL ECG Library

Question 6

In a hospital Holter clinic, a technician finishes a 48-hour Holter hookup on a 50-year-old roofer who plans to work both days and expects to sweat heavily. Which teaching BEST protects the quality of his recording?

  1. Press down or replace any electrode that loosens
  2. Leave the recorder in his truck while he is on the roof
  3. Stay home both days so sweat does not loosen the pads
  4. Unclip the wires on breaks to let his skin dry
Show the answer and rationale

Answer: A

  • A. Correct. Sweat can lift electrodes, and a loose one causes noise or false pauses, so showing him how to press it down or replace it keeps the signal usable.
  • B. Keeping the recorder off his body to protect it seems careful, but it must stay close to his body with the wires connected to keep recording.
  • C. This protects the electrodes but defeats the test, which is meant to show how his heart responds during his usual activities.
  • D. Drying the skin sounds helpful, but every unclipped break leaves a gap in the recording, and reconnecting can loosen the electrodes further.

A Holter recording shows the rhythm during normal daily life, so the patient should keep his usual routine, keep the recorder close to his body and log symptoms in the diary. Good contact is the main threat over 48 hours of work and sweat, because a loose electrode creates baseline wander, noise and false pauses that a reader may mistake for arrhythmia. Patients are told to keep the electrodes firmly attached and are shown how to replace one that loosens, so the technician teaches that step before he leaves.

Sources: LITFL ECG Library

Question 7

At a cardiology office, a 58-year-old man is booked for a 48-hour Holter hookup tomorrow morning. He calls to ask about bathing, because he usually showers every evening. Which instruction should the technician give?

  1. Shower each evening with the recorder sealed in a plastic bag
  2. Shower before the hookup, then keep the recorder dry until it is returned
  3. Take off the electrodes to shower and reapply them on the marked spots
  4. Shower on the second morning, once the first 24 hours are recorded
Show the answer and rationale

Answer: B

  • A. A bag may keep the recorder dry, but the electrodes and wires on the chest still get wet, the adhesive loosens and the tracing fills with artifact or drops out.
  • B. Correct. The recorder and electrodes cannot get wet, so the patient bathes before the appointment and does not shower again until the monitor comes off.
  • C. Removing electrodes on purpose leaves a gap in the continuous record, and electrodes reapplied by the patient may not sit where the leads were planned.
  • D. The ordered test is 48 continuous hours. Halfway is not a safe break, because water reaches the electrodes and recorder just the same on day two.

A Holter monitor records the heart's electrical activity for 24 to 48 hours through adhesive electrodes wired to a small battery-powered recorder worn in a pocket or pouch. Patients cannot bathe or shower while wearing it, so they wash beforehand and keep the device dry until it is returned. The electrodes must stay firmly attached for an accurate recording; if one loosens, the patient follows the instructions the clinic gave at the hookup.

Sources: LITFL ECG Library

Question 8

A technician is fitting a 30-day monitor that records on its own when it senses an irregular rhythm on a 71-year-old man. The order lists the reason as 'stroke, cause unknown; look for atrial fibrillation.' He says, 'I have never felt my heart skip, so this seems like a waste of time.' Which response is BEST?

  1. Explain that pressing its button is what starts a recording, so he should press at any flutter
  2. Tell him it will confirm atrial fibrillation as the cause of his stroke
  3. Agree that having no symptoms is a good sign, but ask him to wear it anyway
  4. Say it can find irregular rhythms people do not feel, and his provider will review the results
Show the answer and rationale

Answer: D

  • A. That describes a patient-activated recorder; this one records by itself, so the explanation leaves him thinking his symptoms decide what is captured.
  • B. No one knows yet whether atrial fibrillation is present, and promising a result is a prediction the technician cannot make.
  • C. Reassuring him sounds kind, but it is misleading here, because the purpose is to find rhythm problems he cannot feel.
  • D. Correct. It gives the reason in plain words, is accurate for a device that triggers by itself, and leaves interpretation to the provider.

Many arrhythmias, including atrial fibrillation, cause no symptoms. After a stroke of unknown cause, a 30-day recording found atrial fibrillation far more often than a 24 to 48 hour Holter in one randomized trial, about 16 percent versus 3 percent for episodes longer than 30 seconds, and the finding changes treatment. A monitor with an automatic trigger stores episodes the patient never feels. The technician explains the purpose in plain words and leaves interpretation and results to the provider.

Sources: LITFL ECG Library

Question 9

At a community hospital's cardiology clinic, a technician finishes fitting a 30-day event monitor on a 72-year-old man. His recordings will go to the monitoring center through a phone app he has never used. Which step BEST makes sure his recordings will reach the center?

  1. Have him send a practice recording to the center before he leaves
  2. Give him the center's phone number to call if a recording will not send
  3. Ask him to bring the monitor in weekly so the clinic can download it
  4. Print the app's steps along with the monitor's serial number
Show the answer and rationale

Answer: A

  • A. Correct. A test sent in the clinic shows that he can complete a transmission and that the center receives a readable tracing, while the technician is there to fix any problem.
  • B. A help number is useful, but he would learn that sending fails only after a real episode, when that recording may already be delayed or lost.
  • C. Weekly visits delay review by days, which defeats sending recordings to a center that can act on serious rhythms, and they add trips for the patient.
  • D. Written steps help him later, but reading them does not prove he can finish a transmission or that the signal arrives legibly.

Event monitors save recordings around symptoms, or automatically in some models, and send them to a monitoring center by telephone or over a cellular link. Staff there review each recording and notify the provider by preset criteria. Failed transmission is a known weak point: in one study of 495 children, more than half never sent a legible ECG while symptomatic. A practice transmission at the hookup, with the technician watching, finds problems while they can still be fixed.

Sources: LITFL ECG Library

Question 10

In a family medicine office, a technician is about to place the electrodes for a 30-day external loop monitor on a 46-year-old woman. While her chest is being prepped, she mentions that surgical tape left her with itchy welts and blisters after an operation last year. Which action should the technician take FIRST?

  1. Apply the electrodes and have her take off any that start to itch
  2. Place one electrode on her forearm and check it in an hour
  3. Stop and tell the ordering provider about her reaction to tape
  4. Set the monitor for 14 days instead of 30 so her skin has less exposure
Show the answer and rationale

Answer: C

  • A. This keeps the test on schedule, but a reaction like hers can recur and worsen over weeks of wear, and electrodes she pulls off leave gaps in the recording.
  • B. A trial patch sounds cautious, but it is a test nobody ordered, and an hour of wear says little about how her skin will handle a month of electrodes.
  • C. Correct. A past blistering reaction to adhesive matters before a month of electrode wear, so the provider hears about it first and decides on the electrodes or the device.
  • D. Less wear time sounds gentler, but the provider chose 30 days to catch infrequent symptoms, and changing the length of an ordered study is not the technician's decision.

Patients are advised to tell their provider about any allergy to tape or other adhesives before an event monitor is placed, because the electrode patches can irritate the skin and skin irritation is the most common problem with these devices. An external loop monitor can be worn for up to 30 days, so a reaction that blisters could spread and cut the study short. The technician stops before any electrode goes on, reports the history, documents it, and follows the provider's direction, which may be different electrodes or a different device.

Sources: LITFL ECG Library; NHA CET Test Plan (2026 job analysis)