# Paper, Settings and Measurement

The grid, paper speed, gain and the calibration pulse, filters, heart rate methods, PR, QRS and QT in adults and children, QTc formulas and U waves

Reviewed by [Alex G, BSN, RN, CEN, CCRN, CMC, NREMT-P](https://icubasics.com/about.html) · Checked against current sources November 2026

[Practice 50 Paper, Settings and Measurement questions free](https://ekgtechbasics.com/app/?from=/sheets/ekg-paper-measurement.html#/topic/ekg-paper-measurement)

[![Page 1 of the Paper, Settings and Measurement PDF, a printed page in columns with sections on The grid at standard settings, Speed, gain and the calibration pulse, Filters, Heart rate, Intervals: normal values, Measuring the QT and QTc and U waves](https://ekgtechbasics.com/sheets/thumbs/ekg-paper-measurement.webp) Enlarge](https://ekgtechbasics.com/sheets/thumbs/ekg-paper-measurement-lg.webp)

**One-page PDF** Letter landscape, ready to print or keep with your notes

[View PDF](https://ekgtechbasics.com/sheets/ekg-paper-measurement.pdf) [Download](https://ekgtechbasics.com/sheets/ekg-paper-measurement.pdf)

## The grid at standard settings

*Horizontal is time, vertical is voltage. A standard 10-second strip at 25 mm/s holds 50 large boxes; at 50 mm/s the same paper holds 5 seconds.*

| Box | Time at 25 mm/s | Time at 50 mm/s | Voltage at 10 mm/mV |
| --- | --- | --- | --- |
| Small (1 mm) | 0.04 s (40 ms) | 0.02 s (20 ms) | 0.1 mV |
| Large (5 mm) | 0.20 s (200 ms) | 0.10 s (100 ms) | 0.5 mV |
| 5 large boxes | 1 second | 0.5 second | 2.5 mV |

## Speed, gain and the calibration pulse

| Setting | What it changes | Check |
| --- | --- | --- |
| Speed 25 mm/s (standard) | Width only | Printed in the header |
| Speed 50 mm/s | Everything twice as wide; intervals look twice as long | Halve box times: small 0.02 s. Rate constants become 600 and 3,000 |
| Gain 10 mm/mV (standard) | Height only | 1 mV calibration pulse is 10 mm tall |
| Gain 5 mm/mV (half standard) | Complexes half as tall | Pulse 5 mm tall; double every amplitude you measure |
| Gain 20 mm/mV (double standard) | Complexes twice as tall | Pulse 20 mm tall; halve every amplitude you measure |

**Exam tip** The calibration pulse is the ruler for height. Read its height before any amplitude: a 10 mm R wave is 1 mV at standard gain, 2 mV at half standard and 0.5 mV at double standard. Paper speed never changes height, and gain never changes width.

Try one

Paper, Settings and Measurement practice question 1 of 3

In a stress lab, a technician prints a resting 12-lead and then sees that the header reads 50 mm/s and 5 mm/mV, settings left by the previous user. Compared with a tracing of the same patient at standard settings, how do the complexes on this printout look?

A. Twice as wide and twice as tall
B. Twice as wide and half as tall
C. Half as wide and half as tall
D. Half as wide and twice as tall

**Show the answer and rationale**

**Answer: B**

- **A.** The speed effect is right, but reading 5 mm/mV as a higher sensitivity reverses the gain: fewer millimeters per millivolt means shorter waves, not taller ones.
- **B.** Correct. Double paper speed spreads each wave over twice the distance, and half the standard gain prints each millivolt as 5 mm instead of 10 mm.
- **C.** The gain effect is right, but faster paper stretches the waves; narrower waves would come from a slower speed such as 12.5 mm/s.
- **D.** This reverses both settings and describes a slower paper speed with a higher gain, the opposite of what the header shows.

Paper speed sets the horizontal scale and gain sets the vertical scale, and each works on its own. At the standard 25 mm/s a small box is 0.04 s; at 50 mm/s the same interval covers twice as many millimeters, so PR, QRS and R-R distances all look twice as long, and a rate counted with the standard box rules comes out at half the true rate. At the standard 10 mm/mV a 1 mV signal prints 10 mm tall; at 5 mm/mV it prints 5 mm, so every wave looks half its usual height and normal voltage can be misread as low. The technician resets both settings to standard, repeats the tracing and confirms a 10 mm calibration mark before passing it on.

Sources: [LITFL ECG Library](https://litfl.com/ecg-library/); [Green, 12-Lead EKG Confidence, 3rd ed.](https://www.aacn.org/-/media/aacn-website/certification/get-certified/handbooks/cmcexamhandbook.pdf)

[Practice 49 more questions free](https://ekgtechbasics.com/app/?from=/sheets/ekg-paper-measurement.html#/topic/ekg-paper-measurement) Rationales for every answer. No download needed.

## Filters

| Mode | Band | Use | Cost |
| --- | --- | --- | --- |
| Diagnostic, adult | 0.05–150 Hz | Every resting 12-lead | Shows noise if the prep is poor |
| Diagnostic, child | 0.05–250 Hz | Pediatric 12-lead | As above |
| Monitoring (telemetry, bedside) | Narrower, about 0.5–40 Hz | Rate and rhythm | ST level unreliable; smaller QRS, lost notches and pacing spikes |

**Safety** A 0.5 Hz low-frequency cutoff, used in real time on monitors, can create false ST elevation or depression of several millimeters. Do not judge ST changes on a monitoring-mode strip; record a 12-lead at diagnostic settings and report to the nurse or provider. Fix noise at its source (skin prep, electrode contact, cables, power cords), not with a filter.

## Heart rate

| Method | How (25 mm/s) | Best for | Fails when |
| --- | --- | --- | --- |
| 300 method | 300 ÷ large boxes R to R (300, 150, 100, 75, 60, 50) | Quick check, regular rhythm | Irregular rhythm; R waves between lines (round with care) |
| 1,500 method | 1,500 ÷ small boxes R to R | Regular rhythm, especially fast | Irregular rhythm |
| 6-second method | QRS complexes in 6 s × 10 (or in 10 s × 6) | Irregular or slow rhythms | Imprecise: each complex is worth 10/min (6/min on a 10 s strip) |

**Exam tip** Count atrial and ventricular rates separately when they differ. The machine's rate can be wrong: it may count tall T waves as QRS complexes (doubling the rate) or miss small complexes. At 50 mm/s use 600 ÷ large boxes or 3,000 ÷ small boxes.

## Intervals: normal values

| Measure | From → to | Adult normal | Children |
| --- | --- | --- | --- |
| Heart rate | R to R | 60–100/min (exam convention) | Newborn 110–150, age 2: 85–125, age 4: 75–115, 6 and over: 60–100 |
| PR interval | Start of P → start of QRS | 0.12–0.20 s (3–5 small boxes) | Shorter: about 0.07–0.15 s in infants, rising with age |
| QRS duration | First to last QRS deflection | 0.07–0.10 s; 0.12 s or more is wide | Narrower: about 0.08 s or less in infants; over 0.09 s may be wide |
| QT interval | Start of QRS → end of T | Under half the R-R at normal rates | Rate-dependent; use QTc |
| QTc | QT corrected for rate | Long: over 450 ms (men), over 460 ms (women); over 500 ms high risk | Up to 440 normal; 440–460 borderline; over 460 prolonged |

## Measuring the QT and QTc

1. Use lead II or V5–V6, whichever shows the longest QT; measure several beats.
2. End of T: draw a line down the steepest part of the T downslope; where it meets the baseline is the end.
3. Leave out a U wave that is separate from the T; if T and U merge, use the tangent.
4. Bazett: QTc = QT ÷ √RR (seconds). Adequate at 60–100/min; it overcorrects above 100 and undercorrects below 60.
5. Outside 60–100/min, Fridericia (QT ÷ ∛RR) or Framingham (QT + 0.154 × (1 − RR)) is preferred; Hodges (QT + 0.00175 × (HR − 60)) is another linear option.

**Exam tip** Worked example at 120/min (RR 0.50 s), QT 0.32 s: Bazett 453 ms looks long, but Fridericia gives 403 ms and Framingham 397 ms. A long Bazett QTc at a fast rate may be the formula, not the patient. Report the QT, the rate and the formula used.

## U waves

A small wave after the T, same direction as the T, usually about 0.5 mm and under a quarter of the T's height, best seen in V2–V3 and at slow rates. Prominent U waves (over 1–2 mm): bradycardia and low potassium most often; also low magnesium or calcium, and drugs such as digoxin, sotalol and amiodarone. An inverted U wave where the T is upright is abnormal: report it.

[Practice Paper, Settings and Measurement questions free in your browser](https://ekgtechbasics.com/app/?from=/sheets/ekg-paper-measurement.html#/topic/ekg-paper-measurement) 50 exam-style questions with rationales. No account or download needed.

## In depth

The one-page summary above is what the PDF prints. These sections go further.

### Why the grid works

ECG paper is a graph. At the standard speed of 25 mm per second, each 1 mm small box is 0.04 s and each 5 mm large box is 0.20 s, so 5 large boxes make 1 second, 300 large boxes make 1 minute, and 1,500 small boxes make 1 minute. That is where the 300 and 1,500 rate constants come from: 300 ÷ large boxes between R waves, or 1,500 ÷ small boxes, gives beats per minute. At the standard gain of 10 mm per millivolt, each small box is 0.1 mV and each large box 0.5 mV. Modern carts print and save PDFs on the same scale.

Electronic calipers report milliseconds. Converting is division: 1,250 ms ÷ 40 ms per small box = 31.25 small boxes, and 1,500 ÷ 31.25 = 48/min; or 60,000 ÷ 1,250 ms = 48/min directly.

At 50 mm/s the paper moves twice as fast, so every wave is twice as wide: a small box is 0.02 s and a large box 0.10 s, and a 10-second paper width holds only 5 seconds. The 300 and 1,500 shortcuts then give half the real rate unless they are doubled to 600 and 3,000. A provider may ask for 50 mm/s to separate closely spaced waves, for example to look for flutter waves or P waves in a fast rhythm. It is not used routinely, because the stretched low-amplitude waves, including U waves, are harder to measure. Height is unchanged by speed: a 2 mm ST deviation is 0.2 mV at either speed if the gain is standard. (At 10 mm/s a small box is 0.1 s and a large box 0.5 s.)

### Gain and the calibration pulse in practice

Most tracings carry a rectangular calibration pulse of 1 mV, 0.2 s wide, usually at the left edge of the tracing or of each row. Its height is the gain actually used: 10 mm at the standard 10 mm/mV, 5 mm at half standard (5 mm/mV) and 20 mm at double standard (20 mm/mV). Half gain fits very tall complexes on the page; double gain enlarges very small ones. Either way, every amplitude on that tracing is read against the pulse, not the grid alone. Voltage criteria such as low QRS voltage (all limb-lead QRS under 5 mm, or all precordial QRS under 10 mm) and the voltage criteria for ventricular hypertrophy assume standard gain.

- Before recording: check the settings screen for 25 mm/s, 10 mm/mV and the diagnostic filter. Settings left by the previous user are a common cause of a misleading tracing.
- After recording: confirm the pulse height matches the gain in the header. A cart that prints a 7 mm pulse at a displayed 10 mm/mV is out of calibration: take it out of use per policy, record on another cart and report it to biomedical engineering.
- Comparing tracings: if today's complexes are half the size of yesterday's, check both pulses before calling it a change in the patient.
- Scanned or faxed copies: if the calibration pulse, speed and gain were cropped off, the reader cannot trust any amplitude; rescan or reprint the full page.
- Mark any non-standard speed or gain clearly so the reader does not measure it as standard.

### Filters: what they remove and what they distort

The skin and electrodes produce a steady offset of up to about 200 mV, far larger than the 3–4 mV heart signal, so every ECG amplifier needs a high-pass (low-frequency cutoff) filter. The question is where to set it. Breathing moves the baseline slowly, at about 0.25 Hz for 15 breaths per minute, and a 0.5 Hz cutoff steadies the baseline. But in a simple real-time filter, a 0.5 Hz cutoff shifts the timing of the low-frequency parts of the ECG, and the error lands where the QRS meets the ST segment. In Buendía-Fuentes 2012 (45 patients), real-time recording at 0.5 Hz produced clinically significant ST shifts in 93%: false elevation in leads with a dominant S wave (V1–V4, inferior leads) and false depression in leads with a dominant R wave (I, aVL, V4–V6), with the J point moved 1.5 to 9 mm (average 3 mm). Wide QRS complexes, such as left bundle branch block and paced beats, were affected most. At 0.05 Hz there was no distortion.

AHA recommendations since 1990 therefore keep 0.05 Hz for traditional filters and allow up to 0.67 Hz only for digital filters that do not shift timing (bidirectional, or linear-phase, filters). Those filters work on stored data, so the computer-processed 12-lead printed a few seconds later can be clean while a real-time printout or monitor at 0.5 Hz is not. That is why ST levels on monitors, telemetry and real-time strips are not used to diagnose ischemia.

The low-pass (high-frequency cutoff) filter removes fast noise: muscle tremor and 50–60 Hz AC interference. Diagnostic adult tracings need at least 150 Hz, and children 250 Hz. A 40 Hz 'muscle' or monitoring filter lowers QRS amplitude and smooths notches and sharp deflections; in one 2026 study, heavy filtering (0.05–25 Hz) cut the QRS notches and slurs seen by 22% and turned about half the notches into slurs, enough for readers to miss a true left bundle branch block pattern. Cutoffs below 150 Hz can also lose pacemaker spikes, so a paced rhythm can be missed. A 40 Hz filter is sometimes used on purpose, for example to clear the interference a left ventricular assist device pump causes, but then only for rhythm, not for detailed QRS or ST-T review.

| Situation | What the technician does |
| --- | --- |
| Muscle or AC noise on a diagnostic 12-lead | Fix the cause first: a relaxed, comfortable patient; fresh electrodes on prepared skin, over bone rather than muscle for limb leads; cables untangled from power cords; nearby electrical devices moved away if allowed |
| A muscle filter was left on | Turn it off and record at 0.05–150 Hz, especially for a paced patient or a chest pain workup |
| A filter is needed to get any usable tracing | Record it, label the tracing with the filter used, and tell the reader |
| ST change seen on telemetry | Report to the nurse; a diagnostic 12-lead is the test for ST changes |

### Heart rate methods in detail

- 300 method: count large boxes between two R waves and divide 300. The sequence 300, 150, 100, 75, 60, 50 is 300 ÷ 1, 2, 3, 4, 5, 6. An R wave that falls between lines makes this an estimate: 3.6 large boxes is 83/min, not 75 or 100.
- 1,500 method: count small boxes and divide 1,500. More precise, and the method of choice for fast regular rhythms, where one box changes the answer by a lot (10 small boxes is 150/min, 9 is 167/min).
- 6-second method: count QRS complexes in 6 seconds (30 large boxes) and multiply by 10, or count a full 10-second strip and multiply by 6. It gives the average rate and is the only one of the three that is valid when the rhythm is irregular.
- Slow rhythms: the count methods are coarse (one complex more or less changes the answer by 6 or 10), so count the whole 10 seconds or measure R to R if the rhythm is regular.
- Two rates: when P waves and QRS complexes are not linked one to one, calculate the atrial rate from P to P and the ventricular rate from R to R and report both.
- Machine rate: usually right, but check it against the strip when T waves are tall, the QRS is small, or there is artifact.

### Measuring PR, QRS, ST and QT

| Measure | Landmarks | Pitfalls |
| --- | --- | --- |
| PR interval | Start of the P wave to the start of the QRS (Q or R, whichever comes first) | Under 0.12 s: possible pre-excitation or junctional origin. Over 0.20 s: first-degree AV block. A slurred QRS start (delta wave) with a short PR needs reporting |
| QRS duration | First deflection of the QRS to the J point (where the QRS ends and the ST begins) | 0.10–0.12 s is abnormal but below bundle branch block width; 0.12 s or more: bundle branch block, pre-excitation, paced or ventricular beats, high potassium, sodium-channel drugs |
| ST level | At the J point, against the baseline set by the PR (or TP) segment | Only on a diagnostic-filter tracing at standard gain |
| QT interval | Start of the QRS to the end of the T | Flat T tail, merged U wave, P wave on the T at fast rates, irregular R-R |

- Lead choice for the QT: lead II or V5–V6, using whichever is longest. Bazett used lead II, and the U wave is usually smallest there; mid-chest leads can overestimate. Use the same lead each time when the QT is being trended.
- Several beats: QT varies by more than 20 ms within one 10-second tracing, and it is longer during sleep.
- Irregular rhythms such as atrial fibrillation or marked sinus arrhythmia: measure several beats and average them.
- Premature beats: do not measure the beat right after a premature beat, because its repolarization is altered.
- Flat or long T tail: use the tangent method; following the tail to the baseline overmeasures the QT.
- U waves and the QT: a small U wave, or one separate from the T, is always left out. Sources differ on a large U wave fused to the T: LITFL includes a fused U wave over 1 mm in the QT, while Lester 2019 and the Brazilian guideline draw the tangent down the T downslope, which ends the measurement before the U. Follow the lab's rule and say which was used. In low potassium, the Brazilian guideline suggests measuring in aVL, where the U wave tends to be flattest.
- Wide QRS: a bundle branch block or paced rhythm lengthens the QT without a true repolarization problem. Correction methods exist (for example subtracting the extra QRS width); record the QRS width beside the QT so the reader can judge.
- Quick screen: at normal rates a QT under half the preceding R-R is normal; over half means calculate the QTc.

### QTc formulas compared

The QT shortens as the rate rises, so it is corrected to what it would be at 60/min (RR 1.0 s). Every formula agrees at exactly 60/min, where QTc equals QT. RR = 60 ÷ heart rate, in seconds.

*Arithmetic examples. At the fast rate Bazett gives the longest value (overcorrection); at the slow rate it gives the shortest (undercorrection).*

| Formula | Equation (seconds) | At 120/min, QT 0.32 s | At 50/min, QT 0.46 s |
| --- | --- | --- | --- |
| Bazett (1920) | QT ÷ √RR | 453 ms | 420 ms |
| Fridericia (1920) | QT ÷ ∛RR | 403 ms | 433 ms |
| Framingham | QT + 0.154 × (1 − RR) | 397 ms | 429 ms |
| Hodges | QT + 0.00175 × (HR − 60) | 425 ms | 443 ms |

Bazett is built into most carts because it is simple, and it is adequate between 60 and 100/min (Brazilian guidance puts the upper limit at 90). Outside that range Fridericia or Framingham is recommended, and the FDA no longer routinely asks for Bazett values in drug trials because Fridericia performs better. In one comparison, Bazett detected 54% of prolonged QT intervals and Fridericia 100%. No formula removes the effect of rate entirely, so a QTc from a tachycardic patient should be read with the rate beside it, and repeated when the rate settles if the question matters.

*All agree that over 500 ms carries a high risk of torsades de pointes. A short QTc is under about 340–350 ms.*

| Source | Long QTc, men | Long QTc, women |
| --- | --- | --- |
| AHA/ACCF/HRS 2009 (upper limit of normal) | Over 450 ms | Over 460 ms |
| LITFL | Over 440 ms | Over 460 ms |
| Brazilian guideline 2022 | Over 450 ms | Over 470 ms |
| Drew 2010 (99th percentile) | Over 470 ms | Over 480 ms |

### Children

A child's ECG is read with the same steps but different numbers. Heart rate is higher and falls with age; LITFL lists 60–100/min from age 6, and the Brazilian guideline says rates reach adult values around age 12. The PR is shorter and the QRS narrower than an adult's, and both lengthen with age, so a QRS that is normal for an adult can be wide for an infant. Use the age-specific table on the cart printout or in the lab's reference, not adult cutoffs.

*Ranges from the Brazilian Society of Cardiology 2022 table (based on Davignon). Other reference studies differ by 10–20 ms; sex differences are small.*

| Age | PR (lead II) | QRS |
| --- | --- | --- |
| 1–3 months | 0.07–0.13 s | 0.02–0.08 s |
| 6–12 months | 0.07–0.16 s | 0.03–0.08 s |
| 1–3 years | 0.08–0.15 s | 0.03–0.08 s |
| 5–8 years | 0.09–0.16 s | 0.03–0.08 s |
| 12–16 years | 0.09–0.18 s | 0.04–0.09 s |

- Pre-excitation in a child: a short PR for age with a delta wave; the QRS may be normal or wide for age. The signs can be subtle, sometimes seen only in V3–V4, and can come and go in infants and children.
- QT in children: measure in II, V5 and V6 and use the longest. At fast rates the next P wave can sit on the T and hide its end. Children often have large U waves: leave out a separate U, and use the tangent when the U merges with the T or is more than half its height. Average several beats in marked sinus arrhythmia.
- QTc in children: up to 440 ms is normal, 440–460 ms borderline, over 460 ms prolonged. Bazett can overcorrect at high pediatric heart rates, although one study in children under 2 found it the most consistent formula; borderline values are confirmed on repeat tracings.
- Filters: the diagnostic band reaches 250 Hz for children.
- Computer interpretation is less accurate in children; the tracing is read by the provider.
- T waves negative in V1–V4 are normal until about age 12.

### Drugs and electrolytes on the measurements

| Cause | What changes on the tracing |
| --- | --- |
| Low potassium | Prominent U waves, ST depression, flat T waves; the T and U can merge into a long 'QU' interval that is easily measured as a long QT |
| High potassium | Tall, peaked T waves first (changes usually appear from about 6.0 mmol/L); then flat, wide P waves and a longer PR, then a wide QRS. The level and the tracing correlate poorly |
| Low magnesium | Long QT; prominent U waves |
| Low calcium | Long, flat ST segment, so a long QT |
| High calcium | Short ST segment, short QT |
| Digoxin | Short QT, ST depression, sometimes prominent U waves |
| Sodium-channel blocking drugs (for example tricyclic overdose) | Wide QRS |
| Class Ia and III antiarrhythmics (quinidine, procainamide, sotalol, amiodarone) and many other drugs | Long QT, prominent U waves |
| Hypothermia | Long QT, wide QRS, prominent U waves |

**Safety** Report promptly to the nurse or provider, per protocol: a QTc over 500 ms or a clear rise from the patient's baseline, a new wide QRS, a QT that keeps lengthening on a QT-prolonging drug, inverted U waves, or any measurement the machine and your own count disagree on. Give the measured QT, the rate, the lead and the formula.

[Practice free in your browser](https://ekgtechbasics.com/app/?from=/sheets/ekg-paper-measurement.html) 650+ exam-style questions with rationales. Free today, no account, no ads. Pro is coming in January 2027. Also free [on Google Play](https://play.google.com/store/apps/details?id=com.alex.ekgtechbasics).

## Related cheat sheets

- [**Safety, Compliance and Patient Care** Patient ID, consent, HIPAA, infection control, electrical safety, vital signs, recognizing and reporting emergencies, CPR and AED, safe lifting, scope and charting for EKG technicians](https://ekgtechbasics.com/sheets/ekg-safety-patient-care.html)
- [**Electrode and Lead Placement** Limb and chest landmarks, counting from the angle of Louis, telemetry and Mason-Likar layouts, right-sided, posterior and 15-lead recordings, children, amputations, dextrocardia, and what to write when placement is not standard](https://ekgtechbasics.com/sheets/ekg-lead-placement.html)
- [**Artifact and Lead Problems** Tremor, 60-cycle interference, wandering baseline, loose electrodes, lead-off and cable faults, artifact that mimics VT, VF or a pause, and limb and chest lead reversals](https://ekgtechbasics.com/sheets/ekg-artifact-troubleshooting.html)
- [**Analyzing a Rhythm Strip Step by Step** A fixed order for every strip, regularity, narrow vs wide fast rhythms, early vs escape beats, describing what you cannot name, and what to report at once](https://ekgtechbasics.com/sheets/ekg-rhythm-analysis-steps.html)
- [**AV Blocks, Bundle Branch Blocks and Pre-excitation** Criteria tables for every degree of AV block, AV dissociation, bundle branch and fascicular blocks, and WPW, with what to measure and when to report](https://ekgtechbasics.com/sheets/ekg-av-blocks-conduction.html)
- [**Reading a 12-Lead** Axis, atrial enlargement, LVH and RVH voltage, ST elevation by lead group, reciprocal change, Q waves, posterior MI, STEMI mimics, and what to report at once](https://ekgtechbasics.com/sheets/ekg-twelve-lead-interpretation.html)

## Sources

1. Life in the Fast Lane (LITFL) ECG Library. ECG Rate Interpretation. Accessed November 2026. [Link](https://litfl.com/ecg-rate-interpretation/)
2. Life in the Fast Lane (LITFL) ECG Library. QT Interval. Accessed November 2026. [Link](https://litfl.com/qt-interval-ecg-library/)
3. Life in the Fast Lane (LITFL) ECG Library. PR Interval. Accessed November 2026. [Link](https://litfl.com/pr-interval-ecg-library/)
4. Life in the Fast Lane (LITFL) ECG Library. QRS Interval. Accessed November 2026. [Link](https://litfl.com/qrs-interval-ecg-library/)
5. Life in the Fast Lane (LITFL) ECG Library. U Wave. Accessed November 2026. [Link](https://litfl.com/u-wave-ecg-library/)
6. Life in the Fast Lane (LITFL) ECG Library. ST Segment. Accessed November 2026. [Link](https://litfl.com/st-segment-ecg-library/)
7. Life in the Fast Lane (LITFL) ECG Library. Low QRS Voltage. Accessed November 2026. [Link](https://litfl.com/low-qrs-voltage-ecg-library/)
8. Life in the Fast Lane (LITFL) ECG Library. Hyperkalaemia. Accessed November 2026. [Link](https://litfl.com/hyperkalaemia-ecg-library/)
9. Cascino T, Shea MJ. Electrocardiography. Merck Manual Professional Edition. Full review May 2026. [Link](https://www.merckmanuals.com/professional/cardiovascular-disorders/cardiovascular-tests-and-procedures/electrocardiography)
10. Buendía-Fuentes F, Arnau-Vives MA, Arnau-Vives A, et al. High-Bandpass Filters in Electrocardiography: Source of Error in the Interpretation of the ST Segment. ISRN Cardiol. 2012;2012:706217. doi:10.5402/2012/706217 [Link](https://europepmc.org/article/PMC/PMC3388307)
11. Gulizia MM, Casolo G, Zuin G, et al. ANMCO/AIIC/SIT Consensus Information Document: definition, precision, and suitability of electrocardiographic signals of electrocardiographs, ergometry, Holter electrocardiogram, telemetry, and bedside monitoring systems. Eur Heart J Suppl. 2017;19(Suppl D):D190-D211. doi:10.1093/eurheartj/sux031 [Link](https://europepmc.org/article/PMC/PMC5520765)
12. Samesima N, God EG, Kruse JCL, et al. Brazilian Society of Cardiology Guidelines on the Analysis and Issuance of Electrocardiographic Reports - 2022. Arq Bras Cardiol. 2022;119(4):638-680. doi:10.36660/abc.20220623 [Link](https://europepmc.org/article/PMC/PMC9563889)
13. Shivashankara KK, Deepanshi, Mehri Shervedani A, Clifford GD, Reyna MA, Sameni R. ECG-Image-Kit: a synthetic image generation toolbox to facilitate deep learning-based electrocardiogram digitization. Physiol Meas. 2024;45(5). doi:10.1088/1361-6579/ad4954 [Link](https://europepmc.org/article/PMC/PMC11135178)
14. Lester RM, Paglialunga S, Johnson IA. QT Assessment in Early Drug Development: The Long and the Short of It. Int J Mol Sci. 2019;20(6):1324. doi:10.3390/ijms20061324 [Link](https://europepmc.org/article/PMC/PMC6471571)
15. Lester RM, Paglialunga S. Surfing the T Wave: A Primer on ECG T Wave Morphologies Encountered in Clinical Trials and Impact on the QT Interval and Patient Safety. Clin Transl Sci. 2025;18(3):e70145. doi:10.1111/cts.70145 [Link](https://europepmc.org/article/PMC/PMC11929567)
16. Miliaraki M, Germanakis I. Pediatric Electrocardiogram in Preparticipation Screening: Narrative Review of Normal Values in Key Features. Children (Basel). 2026;13(2):209. doi:10.3390/children13020209 [Link](https://europepmc.org/article/PMC/PMC12938955)
17. van der Heiden C, Jahangiri P, Caliskan K. HeartMate 3-related ECG artefact resolved by 40-Hz low-pass filtering. Intensive Care Med. 2026;52(4):827-829. doi:10.1007/s00134-026-08347-z [Link](https://europepmc.org/article/PMC/PMC13221310)
18. Zhao Z, Poviser L, Smisek R, Curila K. Inappropriate surface ECG signal filtering significantly reduces physicians' ability to recognize left bundle branch block (abstract). Europace. 2026;28(Suppl 1):euag105.020. doi:10.1093/europace/euag105.020 [Link](https://europepmc.org/article/PMC/PMC13300843)
19. Kligfield P, Gettes LS, Bailey JJ, et al. Recommendations for the standardization and interpretation of the electrocardiogram: part I: the electrocardiogram and its technology. Circulation. 2007;115(10):1306-1324. doi:10.1161/CIRCULATIONAHA.106.180200 (abstract read; filter wording as quoted by Buendía-Fuentes 2012 and Gulizia 2017) [Link](https://europepmc.org/article/MED/17322457)
20. National Healthcareer Association. Certified EKG Technician (CET) Test Plan, based on the 2026 job analysis. [Link](https://www.nhanow.com/docs/default-source/test-plans/nha_2026-cet-test-plan.pdf)
21. National Healthcareer Association. Certified EKG Technician (CET) Test Plan, based on the 2017 job analysis. [Link](https://info.nhanow.com/hubfs/Test%20Plans/nha_2017-cet-test-plan.pdf)