Cheat Sheet

Paper, Settings and Measurement

The grid, paper speed, gain and the calibration pulse, filters, heart rate methods, PR, QRS and QT in adults and children, QTc formulas and U waves

Reviewed by Alex G, BSN, RN, CEN, CCRN, CMC, NREMT-P · Checked against current sources November 2026

Practice 50 Paper, Settings and Measurement questions free

Page 1 of the Paper, Settings and Measurement PDF, a printed page in columns with sections on The grid at standard settings, Speed, gain and the calibration pulse, Filters, Heart rate, Intervals: normal values, Measuring the QT and QTc and U wavesEnlarge

One-page PDFLetter landscape, ready to print or keep with your notes

View PDFDownload

Paper, Settings and Measurement: one-page PDF preview

Page 1 of the Paper, Settings and Measurement PDF, a printed page in columns with sections on The grid at standard settings, Speed, gain and the calibration pulse, Filters, Heart rate, Intervals: normal values, Measuring the QT and QTc and U waves

View PDFDownload

The grid at standard settings

Horizontal is time, vertical is voltage. A standard 10-second strip at 25 mm/s holds 50 large boxes; at 50 mm/s the same paper holds 5 seconds.
BoxTime at 25 mm/sTime at 50 mm/sVoltage at 10 mm/mV
Small (1 mm)0.04 s (40 ms)0.02 s (20 ms)0.1 mV
Large (5 mm)0.20 s (200 ms)0.10 s (100 ms)0.5 mV
5 large boxes1 second0.5 second2.5 mV

Speed, gain and the calibration pulse

SettingWhat it changesCheck
Speed 25 mm/s (standard)Width onlyPrinted in the header
Speed 50 mm/sEverything twice as wide; intervals look twice as longHalve box times: small 0.02 s. Rate constants become 600 and 3,000
Gain 10 mm/mV (standard)Height only1 mV calibration pulse is 10 mm tall
Gain 5 mm/mV (half standard)Complexes half as tallPulse 5 mm tall; double every amplitude you measure
Gain 20 mm/mV (double standard)Complexes twice as tallPulse 20 mm tall; halve every amplitude you measure

Filters

ModeBandUseCost
Diagnostic, adult0.05–150 HzEvery resting 12-leadShows noise if the prep is poor
Diagnostic, child0.05–250 HzPediatric 12-leadAs above
Monitoring (telemetry, bedside)Narrower, about 0.5–40 HzRate and rhythmST level unreliable; smaller QRS, lost notches and pacing spikes

Heart rate

MethodHow (25 mm/s)Best forFails when
300 method300 ÷ large boxes R to R (300, 150, 100, 75, 60, 50)Quick check, regular rhythmIrregular rhythm; R waves between lines (round with care)
1,500 method1,500 ÷ small boxes R to RRegular rhythm, especially fastIrregular rhythm
6-second methodQRS complexes in 6 s × 10 (or in 10 s × 6)Irregular or slow rhythmsImprecise: each complex is worth 10/min (6/min on a 10 s strip)

Intervals: normal values

MeasureFrom → toAdult normalChildren
Heart rateR to R60–100/min (exam convention)Newborn 110–150, age 2: 85–125, age 4: 75–115, 6 and over: 60–100
PR intervalStart of P → start of QRS0.12–0.20 s (3–5 small boxes)Shorter: about 0.07–0.15 s in infants, rising with age
QRS durationFirst to last QRS deflection0.07–0.10 s; 0.12 s or more is wideNarrower: about 0.08 s or less in infants; over 0.09 s may be wide
QT intervalStart of QRS → end of TUnder half the R-R at normal ratesRate-dependent; use QTc
QTcQT corrected for rateLong: over 450 ms (men), over 460 ms (women); over 500 ms high riskUp to 440 normal; 440–460 borderline; over 460 prolonged

Measuring the QT and QTc

  1. Use lead II or V5–V6, whichever shows the longest QT; measure several beats.
  2. End of T: draw a line down the steepest part of the T downslope; where it meets the baseline is the end.
  3. Leave out a U wave that is separate from the T; if T and U merge, use the tangent.
  4. Bazett: QTc = QT ÷ √RR (seconds). Adequate at 60–100/min; it overcorrects above 100 and undercorrects below 60.
  5. Outside 60–100/min, Fridericia (QT ÷ ∛RR) or Framingham (QT + 0.154 × (1 − RR)) is preferred; Hodges (QT + 0.00175 × (HR − 60)) is another linear option.

U waves

A small wave after the T, same direction as the T, usually about 0.5 mm and under a quarter of the T's height, best seen in V2–V3 and at slow rates. Prominent U waves (over 1–2 mm): bradycardia and low potassium most often; also low magnesium or calcium, and drugs such as digoxin, sotalol and amiodarone. An inverted U wave where the T is upright is abnormal: report it.

Practice Paper, Settings and Measurement questions free in your browser50 exam-style questions with rationales. No account or download needed.

In depth

The one-page summary above is what the PDF prints. These sections go further.

Why the grid works

ECG paper is a graph. At the standard speed of 25 mm per second, each 1 mm small box is 0.04 s and each 5 mm large box is 0.20 s, so 5 large boxes make 1 second, 300 large boxes make 1 minute, and 1,500 small boxes make 1 minute. That is where the 300 and 1,500 rate constants come from: 300 ÷ large boxes between R waves, or 1,500 ÷ small boxes, gives beats per minute. At the standard gain of 10 mm per millivolt, each small box is 0.1 mV and each large box 0.5 mV. Modern carts print and save PDFs on the same scale.

Electronic calipers report milliseconds. Converting is division: 1,250 ms ÷ 40 ms per small box = 31.25 small boxes, and 1,500 ÷ 31.25 = 48/min; or 60,000 ÷ 1,250 ms = 48/min directly.

At 50 mm/s the paper moves twice as fast, so every wave is twice as wide: a small box is 0.02 s and a large box 0.10 s, and a 10-second paper width holds only 5 seconds. The 300 and 1,500 shortcuts then give half the real rate unless they are doubled to 600 and 3,000. A provider may ask for 50 mm/s to separate closely spaced waves, for example to look for flutter waves or P waves in a fast rhythm. It is not used routinely, because the stretched low-amplitude waves, including U waves, are harder to measure. Height is unchanged by speed: a 2 mm ST deviation is 0.2 mV at either speed if the gain is standard. (At 10 mm/s a small box is 0.1 s and a large box 0.5 s.)

Gain and the calibration pulse in practice

Most tracings carry a rectangular calibration pulse of 1 mV, 0.2 s wide, usually at the left edge of the tracing or of each row. Its height is the gain actually used: 10 mm at the standard 10 mm/mV, 5 mm at half standard (5 mm/mV) and 20 mm at double standard (20 mm/mV). Half gain fits very tall complexes on the page; double gain enlarges very small ones. Either way, every amplitude on that tracing is read against the pulse, not the grid alone. Voltage criteria such as low QRS voltage (all limb-lead QRS under 5 mm, or all precordial QRS under 10 mm) and the voltage criteria for ventricular hypertrophy assume standard gain.

Filters: what they remove and what they distort

The skin and electrodes produce a steady offset of up to about 200 mV, far larger than the 3–4 mV heart signal, so every ECG amplifier needs a high-pass (low-frequency cutoff) filter. The question is where to set it. Breathing moves the baseline slowly, at about 0.25 Hz for 15 breaths per minute, and a 0.5 Hz cutoff steadies the baseline. But in a simple real-time filter, a 0.5 Hz cutoff shifts the timing of the low-frequency parts of the ECG, and the error lands where the QRS meets the ST segment. In Buendía-Fuentes 2012 (45 patients), real-time recording at 0.5 Hz produced clinically significant ST shifts in 93%: false elevation in leads with a dominant S wave (V1–V4, inferior leads) and false depression in leads with a dominant R wave (I, aVL, V4–V6), with the J point moved 1.5 to 9 mm (average 3 mm). Wide QRS complexes, such as left bundle branch block and paced beats, were affected most. At 0.05 Hz there was no distortion.

AHA recommendations since 1990 therefore keep 0.05 Hz for traditional filters and allow up to 0.67 Hz only for digital filters that do not shift timing (bidirectional, or linear-phase, filters). Those filters work on stored data, so the computer-processed 12-lead printed a few seconds later can be clean while a real-time printout or monitor at 0.5 Hz is not. That is why ST levels on monitors, telemetry and real-time strips are not used to diagnose ischemia.

The low-pass (high-frequency cutoff) filter removes fast noise: muscle tremor and 50–60 Hz AC interference. Diagnostic adult tracings need at least 150 Hz, and children 250 Hz. A 40 Hz 'muscle' or monitoring filter lowers QRS amplitude and smooths notches and sharp deflections; in one 2026 study, heavy filtering (0.05–25 Hz) cut the QRS notches and slurs seen by 22% and turned about half the notches into slurs, enough for readers to miss a true left bundle branch block pattern. Cutoffs below 150 Hz can also lose pacemaker spikes, so a paced rhythm can be missed. A 40 Hz filter is sometimes used on purpose, for example to clear the interference a left ventricular assist device pump causes, but then only for rhythm, not for detailed QRS or ST-T review.

SituationWhat the technician does
Muscle or AC noise on a diagnostic 12-leadFix the cause first: a relaxed, comfortable patient; fresh electrodes on prepared skin, over bone rather than muscle for limb leads; cables untangled from power cords; nearby electrical devices moved away if allowed
A muscle filter was left onTurn it off and record at 0.05–150 Hz, especially for a paced patient or a chest pain workup
A filter is needed to get any usable tracingRecord it, label the tracing with the filter used, and tell the reader
ST change seen on telemetryReport to the nurse; a diagnostic 12-lead is the test for ST changes

Heart rate methods in detail

Measuring PR, QRS, ST and QT

MeasureLandmarksPitfalls
PR intervalStart of the P wave to the start of the QRS (Q or R, whichever comes first)Under 0.12 s: possible pre-excitation or junctional origin. Over 0.20 s: first-degree AV block. A slurred QRS start (delta wave) with a short PR needs reporting
QRS durationFirst deflection of the QRS to the J point (where the QRS ends and the ST begins)0.10–0.12 s is abnormal but below bundle branch block width; 0.12 s or more: bundle branch block, pre-excitation, paced or ventricular beats, high potassium, sodium-channel drugs
ST levelAt the J point, against the baseline set by the PR (or TP) segmentOnly on a diagnostic-filter tracing at standard gain
QT intervalStart of the QRS to the end of the TFlat T tail, merged U wave, P wave on the T at fast rates, irregular R-R

QTc formulas compared

The QT shortens as the rate rises, so it is corrected to what it would be at 60/min (RR 1.0 s). Every formula agrees at exactly 60/min, where QTc equals QT. RR = 60 ÷ heart rate, in seconds.

Arithmetic examples. At the fast rate Bazett gives the longest value (overcorrection); at the slow rate it gives the shortest (undercorrection).
FormulaEquation (seconds)At 120/min, QT 0.32 sAt 50/min, QT 0.46 s
Bazett (1920)QT ÷ √RR453 ms420 ms
Fridericia (1920)QT ÷ ∛RR403 ms433 ms
FraminghamQT + 0.154 × (1 − RR)397 ms429 ms
HodgesQT + 0.00175 × (HR − 60)425 ms443 ms

Bazett is built into most carts because it is simple, and it is adequate between 60 and 100/min (Brazilian guidance puts the upper limit at 90). Outside that range Fridericia or Framingham is recommended, and the FDA no longer routinely asks for Bazett values in drug trials because Fridericia performs better. In one comparison, Bazett detected 54% of prolonged QT intervals and Fridericia 100%. No formula removes the effect of rate entirely, so a QTc from a tachycardic patient should be read with the rate beside it, and repeated when the rate settles if the question matters.

All agree that over 500 ms carries a high risk of torsades de pointes. A short QTc is under about 340–350 ms.
SourceLong QTc, menLong QTc, women
AHA/ACCF/HRS 2009 (upper limit of normal)Over 450 msOver 460 ms
LITFLOver 440 msOver 460 ms
Brazilian guideline 2022Over 450 msOver 470 ms
Drew 2010 (99th percentile)Over 470 msOver 480 ms

Children

A child's ECG is read with the same steps but different numbers. Heart rate is higher and falls with age; LITFL lists 60–100/min from age 6, and the Brazilian guideline says rates reach adult values around age 12. The PR is shorter and the QRS narrower than an adult's, and both lengthen with age, so a QRS that is normal for an adult can be wide for an infant. Use the age-specific table on the cart printout or in the lab's reference, not adult cutoffs.

Ranges from the Brazilian Society of Cardiology 2022 table (based on Davignon). Other reference studies differ by 10–20 ms; sex differences are small.
AgePR (lead II)QRS
1–3 months0.07–0.13 s0.02–0.08 s
6–12 months0.07–0.16 s0.03–0.08 s
1–3 years0.08–0.15 s0.03–0.08 s
5–8 years0.09–0.16 s0.03–0.08 s
12–16 years0.09–0.18 s0.04–0.09 s

Drugs and electrolytes on the measurements

CauseWhat changes on the tracing
Low potassiumProminent U waves, ST depression, flat T waves; the T and U can merge into a long 'QU' interval that is easily measured as a long QT
High potassiumTall, peaked T waves first (changes usually appear from about 6.0 mmol/L); then flat, wide P waves and a longer PR, then a wide QRS. The level and the tracing correlate poorly
Low magnesiumLong QT; prominent U waves
Low calciumLong, flat ST segment, so a long QT
High calciumShort ST segment, short QT
DigoxinShort QT, ST depression, sometimes prominent U waves
Sodium-channel blocking drugs (for example tricyclic overdose)Wide QRS
Class Ia and III antiarrhythmics (quinidine, procainamide, sotalol, amiodarone) and many other drugsLong QT, prominent U waves
HypothermiaLong QT, wide QRS, prominent U waves
Practice free in your browser650+ exam-style questions with rationales. Free today, no account, no ads. Pro is coming in January 2027. Also free on Google Play.

Related cheat sheets

Sources

  1. Life in the Fast Lane (LITFL) ECG Library. ECG Rate Interpretation. Accessed November 2026. Link
  2. Life in the Fast Lane (LITFL) ECG Library. QT Interval. Accessed November 2026. Link
  3. Life in the Fast Lane (LITFL) ECG Library. PR Interval. Accessed November 2026. Link
  4. Life in the Fast Lane (LITFL) ECG Library. QRS Interval. Accessed November 2026. Link
  5. Life in the Fast Lane (LITFL) ECG Library. U Wave. Accessed November 2026. Link
  6. Life in the Fast Lane (LITFL) ECG Library. ST Segment. Accessed November 2026. Link
  7. Life in the Fast Lane (LITFL) ECG Library. Low QRS Voltage. Accessed November 2026. Link
  8. Life in the Fast Lane (LITFL) ECG Library. Hyperkalaemia. Accessed November 2026. Link
  9. Cascino T, Shea MJ. Electrocardiography. Merck Manual Professional Edition. Full review May 2026. Link
  10. Buendía-Fuentes F, Arnau-Vives MA, Arnau-Vives A, et al. High-Bandpass Filters in Electrocardiography: Source of Error in the Interpretation of the ST Segment. ISRN Cardiol. 2012;2012:706217. doi:10.5402/2012/706217 Link
  11. Gulizia MM, Casolo G, Zuin G, et al. ANMCO/AIIC/SIT Consensus Information Document: definition, precision, and suitability of electrocardiographic signals of electrocardiographs, ergometry, Holter electrocardiogram, telemetry, and bedside monitoring systems. Eur Heart J Suppl. 2017;19(Suppl D):D190-D211. doi:10.1093/eurheartj/sux031 Link
  12. Samesima N, God EG, Kruse JCL, et al. Brazilian Society of Cardiology Guidelines on the Analysis and Issuance of Electrocardiographic Reports - 2022. Arq Bras Cardiol. 2022;119(4):638-680. doi:10.36660/abc.20220623 Link
  13. Shivashankara KK, Deepanshi, Mehri Shervedani A, Clifford GD, Reyna MA, Sameni R. ECG-Image-Kit: a synthetic image generation toolbox to facilitate deep learning-based electrocardiogram digitization. Physiol Meas. 2024;45(5). doi:10.1088/1361-6579/ad4954 Link
  14. Lester RM, Paglialunga S, Johnson IA. QT Assessment in Early Drug Development: The Long and the Short of It. Int J Mol Sci. 2019;20(6):1324. doi:10.3390/ijms20061324 Link
  15. Lester RM, Paglialunga S. Surfing the T Wave: A Primer on ECG T Wave Morphologies Encountered in Clinical Trials and Impact on the QT Interval and Patient Safety. Clin Transl Sci. 2025;18(3):e70145. doi:10.1111/cts.70145 Link
  16. Miliaraki M, Germanakis I. Pediatric Electrocardiogram in Preparticipation Screening: Narrative Review of Normal Values in Key Features. Children (Basel). 2026;13(2):209. doi:10.3390/children13020209 Link
  17. van der Heiden C, Jahangiri P, Caliskan K. HeartMate 3-related ECG artefact resolved by 40-Hz low-pass filtering. Intensive Care Med. 2026;52(4):827-829. doi:10.1007/s00134-026-08347-z Link
  18. Zhao Z, Poviser L, Smisek R, Curila K. Inappropriate surface ECG signal filtering significantly reduces physicians' ability to recognize left bundle branch block (abstract). Europace. 2026;28(Suppl 1):euag105.020. doi:10.1093/europace/euag105.020 Link
  19. Kligfield P, Gettes LS, Bailey JJ, et al. Recommendations for the standardization and interpretation of the electrocardiogram: part I: the electrocardiogram and its technology. Circulation. 2007;115(10):1306-1324. doi:10.1161/CIRCULATIONAHA.106.180200 (abstract read; filter wording as quoted by Buendía-Fuentes 2012 and Gulizia 2017) Link
  20. National Healthcareer Association. Certified EKG Technician (CET) Test Plan, based on the 2026 job analysis. Link
  21. National Healthcareer Association. Certified EKG Technician (CET) Test Plan, based on the 2017 job analysis. Link