Pacemakers and Implanted Devices on the ECG
The pacemaker code, atrial, ventricular, dual-chamber and biventricular pacing, capture and sensing problems, magnets, ICD shocks, and filters that hide or fake spikes
Practice 18 Pacemakers and Implanted Devices on the ECG questions free
The pacemaker code (NBG)
| Letter | Means | Choices |
|---|---|---|
| 1st | Chamber paced | A atrium, V ventricle, D both |
| 2nd | Chamber sensed | A, V, D, O |
| 3rd | Response to a sensed beat | I inhibit, T trigger, D both, O none |
| 4th | Rate response | R raises the rate with activity, O none |
| 5th | Multisite pacing | For example both ventricles; O none |
VVIR: paces and senses the ventricle, holds back when it senses a beat, speeds up with activity. DDD is the most common mode. AOO, VOO and DOO pace at a fixed rate and sense nothing (asynchronous).
What paced beats look like
| Pacing | On the strip |
|---|---|
| Spike | Narrow vertical line, about 2 ms; tiny with bipolar leads, so check every lead |
| Atrial | Spike, then a P wave; the patient's own QRS follows |
| Ventricular (RV lead) | Spike, then a wide QRS shaped like LBBB, ST-T opposite the QRS |
| Dual-chamber | Spikes before the P only, the QRS only, or both, depending on the patient's own beats |
| Biventricular (CRT) | Paced QRS often upright in V1; two ventricular spikes may show as one |
| Fusion | Spike on a QRS that is a hybrid of paced and native; normal function |
| No spikes | Own rhythm is fast enough: a demand pacemaker waits and senses |
Naming the problem
| Problem | On the strip | Common causes |
|---|---|---|
| Failure to capture | Spike with no P wave or QRS after it | Lead moved or broken, battery, threshold rise; electrolytes, drugs, MI |
| Undersensing (failure to sense) | Spikes ignore the patient's own beats; can land on a T wave | Lead problem, small native signals |
| Oversensing | Pause with no spike when one is due; often with muscle noise | T waves, muscle activity or interference read as beats |
| Failure to pace | No spikes; rate falls below the programmed rate | Oversensing, battery or generator, poor lead contact |
| Paced rate too fast | Paced rate faster than expected, up to the device's upper rate | Tracking a fast atrial rhythm, PMT, activity sensor |
Magnets and ICDs
| Device | Magnet on (stops when it is lifted) |
|---|---|
| Most pacemakers | Fixed-rate pacing at the maker's magnet rate; spikes march through the patient's own beats |
| Transvenous ICD | Shocks and antitachycardia pacing suspended; no fixed-rate pacing |
| Micra leadless pacemaker | No magnet response |
- Placing a magnet is the provider's decision; record and report what the strip shows.
- An ICD also paces, so its spikes look like a pacemaker's. Antitachycardia pacing is a short burst of fast paced beats that can stop VT without a shock.
- Shock, patient awake: stay, tell the nurse, save the strip and the time. Several shocks or symptoms: urgent.
- Shock, patient unresponsive: code team, CPR, AED with the pad off the device. Touching a patient during an ICD shock is not harmful.
Filters, monitors and look-alikes
- Record first at the diagnostic setting (to 150 Hz); a 40 Hz filter can blunt, hide or distort spikes. Label any filter.
- Many monitors find spikes only with their pacer setting on and then draw their own marker; with it off, paced beats can alarm as a ventricular rhythm.
- Fake spikes: static from clothing, loose or broken wires, nerve stimulators (fixed-rate spikes unrelated to the P or QRS).
Exam pearls
In depth
The one-page summary above is what the PDF prints. These sections go further.
Spikes: why they are small, missing or fake
A pacing stimulus is very brief, so on paper it is a thin vertical line, usually about 2 ms wide (LITFL). Its height depends on where the lead sits and what kind of lead it is: bipolar leads make much smaller spikes than unipolar leads, and epicardial leads make smaller spikes than leads inside the heart. A spike that is obvious in one lead may be invisible in the next, which is why CanadiEM advises checking every lead before deciding a beat is not paced.
Machines have to work to see them. In one study of electrocardiographs, pacer spike detection from the ordinary 500-samples-per-second data stream found 77.5% of spikes, while a detector sampling at 32,000 per second found 99.5% (Luo 2008). The same study notes that a slow data stream can show only one spike when two ventricular leads fire close together, as in biventricular pacing. Stored ECGs are usually compressed, and SCST warns that compression changes short, sharp signals such as pacemaker spikes more than it changes the ST segment, so an archived copy may not match the original.
Many bedside monitors use a separate pacer detection mode with a higher sampling rate. In Drew's ICU study, when it detected a spike the monitor painted its own marker on the display, so the spike you see may be the monitor's mark, not the real signal (Drew 2014). In that study only one third of patients with pacemakers had the mode switched on, and intermittent ventricular pacing without it produced false 'accelerated ventricular rhythm' alarms. The technician confirms the pacemaker in the chart and asks the nurse to turn the setting on, or turns it on where unit policy allows.
| Filter or setting | Effect on spikes | What to do |
|---|---|---|
| Diagnostic band, 0.05 or 0.67 to 150 Hz (SCST) | Keeps short, sharp signals | Use it for the first recording |
| 40 Hz muscle filter | Distorts the ECG; spikes can be blunted or lost | Only after relaxing the patient fails; label the tracing |
| Any filtering or processing | Spikes can be exaggerated or distorted; tall or broad spikes can pass for fibrillation or noise (Mondal 2026) | Compare with the patient's other tracings; note the device |
| Analog-type monitors | Drop expected spikes more often than diagnostic ECGs (Mondal 2026) | Confirm on a 12-lead or paper strip before calling a malfunction |
Things that make spike-like marks: static electricity, for example from clothing (Luo's test set included spikes from clothing static and a static gun); fractured wires and loose connectors; and other implanted stimulators. Mondal describes neurostimulators (deep brain, sacral nerve and gastric) sending periodic pulses, and deep brain stimulators causing high-frequency artifact, usually 60 to 200 Hz. A stimulator's spikes run at their own fixed rate with no link to the P waves or QRS complexes, and re-prepping the skin does not remove them; ask about implanted devices and tell the nurse or provider. Mondal's rule of thumb: spikes that occur at predictable intervals in relation to atrial or ventricular events are almost always real pacing. Pseudopacing artifacts that no pacing algorithm can explain also occur, and artifact can simulate pacemaker malfunction (ISHNE-HRS), so read every suspected malfunction in context before it is reported as one.
Modes and how a dual-chamber pacemaker behaves
The code follows the NASPE/BPEG (NBG) system, revised in 2002; LITFL notes that many textbooks still print the 1987 version, in which the fourth letter also covered programmability and the fifth covered antitachycardia functions. Positions I to III are the same in both. Merck's table gives position V as O, A, V or D, where multisite pacing means both atria, both ventricles or more than one lead in one chamber; a code can have three to five letters.
| Mode | What it does | What the strip shows |
|---|---|---|
| AAI | Paces and senses the atrium; holds back for the patient's own P wave | Atrial spikes only when the sinus rate falls below the set rate |
| VVI(R) | Paces and senses the ventricle; one of the two most used modes | Ventricular spikes only when the patient's own QRS is late |
| DDD(R) | Senses and paces both chambers; the most common mode | No spikes, atrial spike only, ventricular spike only, or both |
| AOO, VOO, DOO | Paces at a fixed rate and senses nothing; what most pacemakers do under a magnet | Spikes march on regardless of the patient's own beats |
A DDD pacemaker watches the atrium first. If no P wave comes by the time the lower rate is due, it paces the atrium; then it waits a set delay for the patient's own QRS and paces the ventricle only if none comes (Fernando 2026). That is why LITFL says dual-chamber spikes may precede only the P wave, only the QRS, or both. A P wave with a ventricular spike after it means the device is following the patient's own atrial rate; two spikes per beat means both chambers are paced; no spikes means the patient's own rhythm is in charge.
Atrial pacing can unmask a slow AV node. LITFL shows atrially paced patients with first-degree block or Wenckebach conduction that their slower native rhythm did not show, and calls it clinically unimportant if cardiac output holds. Each atrial spike followed by a P wave proves atrial capture; the dropped beats are still reported to the nurse.
Rate response (R) uses a sensor to speed up pacing with activity. Merck notes that such devices can also respond to vibration and muscle activity, so a bumpy car ride can raise the paced rate; a diary entry beside the faster pacing helps the device clinic. Mode switching protects dual-chamber patients from fast atrial rhythms: in atrial fibrillation, for example, a DDDR device can switch itself to VVIR and pace the ventricle at a programmed rate instead of following the atria. The device is working, but new AF is still a rhythm change to report.
Reading a paced strip step by step
- Read the strip as usual, then add four checks (CanadiEM): rate, capture, pacing, sensing.
- Rate: measure spike to spike. At 25 mm/s, 1500 divided by the small boxes between spikes is the paced rate (25 small boxes = 60/min). Compare it with the programmed rate in the chart or on the device card.
- Capture: is every spike followed by its P wave (atrial) or QRS (ventricular)? A spike that lands just after the patient's own beat finds the heart refractory, so no complex follows; that alone does not prove failure to capture. Look for spikes that fall where the heart could respond.
- Pacing: are spikes there whenever the patient's own rate drops below the programmed rate? A pause longer than the pacing interval with no spike is failure to pace, from oversensing or a hardware problem (CanadiEM).
- Sensing: do spikes appear when the patient's own beats are already on time? That is undersensing.
- Record the device, the findings and the time, and report as below. Leave interrogation, reprogramming and treatment to the nurse, provider and device team.
Fusion and pseudofusion are normal behaviour, not malfunction. When the patient's own rate is close to the programmed rate, the device's timer can run out just as a native beat begins. If both wavefronts share the ventricle, the QRS is a hybrid of the paced and native shapes, often narrower than a fully paced beat (fusion, LITFL and CanadiEM). If the native beat already owns the ventricle, the spike sits on a QRS that it did not change (pseudofusion). Check the QRS shape and timing before calling either one a problem.
| Finding | How fast to report |
|---|---|
| Failure to capture or to pace, pauses, or any finding with dizziness, weakness or fainting | Now: stay with the patient and tell the nurse |
| Undersensing, especially a spike on a T wave | Promptly, even without symptoms |
| Paced rate faster or slower than programmed; new AF in a paced patient | Promptly, with the strip |
| Normal paced rhythm, fusion, or no spikes with a normal native rhythm | Document the device; routine |
Malfunctions in more depth
Capture means the stimulus is strong enough to depolarize the heart. CanadiEM groups the causes of failure to capture as hardware (a broken lead or one that has lost contact, battery problems), exit block (a rise in the stimulation threshold where the lead meets the heart), and patient factors (electrolyte problems, drugs, myocardial infarction). Each uncaptured spike is a lost beat, so the true heart rate falls below the programmed rate, which explains weakness or lightheadedness.
Sensing is set by a sensitivity number: a lower number makes the device see smaller signals. Undersensing means the device misses the patient's own beats and paces when it should not, so spikes appear anywhere, from a lead fracture, poor contact, an insulation problem or low-voltage native signals. A spike near the peak of the T wave can cause an R-on-T event, which CanadiEM links to VT or torsades.
Oversensing means the device mistakes something else for a heartbeat and holds back its output: T waves, signals from the wrong chamber, muscle potentials (myopotentials), or electromagnetic interference. The strip shows pauses with no spikes, for example during arm work with muscle noise on the baseline, which separates it from failure to capture (spikes with nothing after them). ISHNE-HRS lists myopotential inhibition, cross-talk and pacemaker-mediated tachycardia as the main abnormalities seen on Holter recordings in pacemaker patients, though they are uncommon, and notes that in one report pacemaker dysfunction explained syncope in only 4.9% of pacemaker patients.
| Problem | Mechanism (Merck, CanadiEM) | On the strip |
|---|---|---|
| Cross-talk | The ventricular channel senses the atrial pacing pulse and withholds ventricular pacing | Atrial spike and P wave, then no ventricular spike and a pause |
| Pacemaker-mediated tachycardia (PMT) | A PVC or paced beat conducts back to the atria; the atrial lead senses that P wave and the device paces the ventricle again, over and over | Sudden ventricular-paced tachycardia at or near the upper rate, often after a PVC, with a retrograde P after each QRS |
| Sensor-driven tachycardia | The rate sensor responds to vibration or muscle activity | Paced rate rises without exertion; match it to the diary |
| Pacemaker syndrome | Ventricular pacing that loses the normal timing between atria and ventricles | Paced rhythm that looks normal; the patient has vague dizziness, neck or breathing symptoms. Report them. |
PMT needs an atrial lead to sense the retrograde P wave, so it occurs only with dual-chamber devices; the device team can stop and prevent it, so the technician reports it promptly. A patient whose own heart produces little or no escape rhythm is called pacing dependent; pauses from oversensing or cross-talk matter most in these patients.
Biventricular pacing (CRT)
Cardiac resynchronization therapy adds a lead that paces the left ventricle, usually from a coronary vein, so both ventricles are paced together. Pacing from the right ventricular apex alone gives a wide QRS that is negative in V1 (recorded at the 4th intercostal space) with, usually, a left superior axis. During biventricular pacing the QRS most often has a dominant R wave in V1 and V2, and the frontal axis most often points to the right superior quadrant, though other axes do not by themselves mean a problem (van Stipdonk 2015).
A negative QRS in V1 during CRT warrants evaluation. Causes include a V1 electrode placed in the wrong space, slow conduction away from the left ventricular lead (which leaves a picture dominated by right ventricular pacing) and a dislodged left ventricular lead. The technician's first job is to confirm that V1 sits in the 4th intercostal space at the right sternal edge, then to report a change in the paced QRS shape compared with earlier tracings.
CRT helps only when it paces nearly every beat. ISHNE-HRS says the benefit depends mostly on effective biventricular pacing, and that programming should aim for the maximum, above 98%. The 2023 HRS/EHRA/APHRS/LAHRS remote device clinic consensus notes that studies have used cut-offs from above 80% to above 98% and names above 95% as the target. PVCs, AF and fusion or pseudofusion beats all lower the true share. In one small study of CRT patients in AF whose devices counted over 90% biventricular pacing, a 12-lead Holter found fusion or pseudofusion beats in 53% of the patients, so a high device count can overstate real pacing. A high PVC or AF burden on a Holter is a finding the device clinic needs.
Magnet response in more depth
Most pacemakers switch to asynchronous pacing (AOO, VOO or DOO, depending on the leads) while a magnet is over them, at a rate set by the maker. Sensing is off, so spikes compete with the patient's own beats, and spikes that land just after a native beat find the ventricle refractory; that is expected while the magnet is on. LITFL notes that asynchronous ventricular pacing carries a risk of pacemaker-induced VT, so a spike on a T wave is still reported. The magnet rate is lower once the battery reaches its elective replacement point (ERI), by about 20% (Fernando 2026).
| Maker (Fernando 2026, general responses) | Normal magnet rate | At replacement (ERI) |
|---|---|---|
| Abbott (includes St. Jude) | 100/min | 85/min |
| Biotronik | 90/min | 80/min |
| Boston Scientific | 100/min | 85/min |
| Medtronic | 85/min | 65/min |
| Microport | 96/min | 80/min |
- Transvenous ICD: a magnet suspends shocks and antitachycardia pacing in every brand; no current transvenous ICD switches to asynchronous pacing. Some older Microport (Sorin) ICDs raise their pacing rate under a magnet (96/min, falling to 80/min near replacement) without changing mode, which does not mean the device is a pacemaker.
- Subcutaneous ICD: no leads inside the heart; a magnet suspends shocks and usually produces a tone for about 60 seconds.
- Leadless pacemakers: the Micra does not respond to a magnet; the AVEIR switches to asynchronous pacing at 100/min, lower as the battery runs down.
- Removing the magnet restores the device's programmed behaviour.
- Magnets and MRI are outside the technician's decisions: if a patient with a device is booked for MRI, or a magnet is being used, record what the strip shows and report it.
ICDs: pacing, antitachycardia pacing and shocks
A transvenous ICD has leads in the right ventricle and sometimes the right atrium (a CRT-D adds a left ventricular lead). It cardioverts or defibrillates VT and VF, and tiered-therapy ICDs also pace for slow rhythms and use antitachycardia pacing (ATP) on tachycardias that respond to it; they store intracardiac electrograms of each event (Merck). Its bradycardia pacing looks the same on the ECG as a pacemaker's.
ATP delivers a burst of stimuli slightly faster than the tachycardia; in Okubo's cases the burst interval was 88% of the VT cycle length. On a monitor or Holter it appears as a short run of rapid paced beats interrupting the VT, followed by the return of the patient's rhythm if it worked. It is painless, but it fails more often in fast VT (success as low as 50% when the cycle length is under 300 ms), and it can accelerate VT in up to 10% of therapies, after which the device may shock.
Merck lists the causes of inappropriate shocks or pacing: sinus rhythm, supraventricular tachycardia, atrial fibrillation or flutter, and nonphysiologic signals such as those from a fractured lead. Missed therapy can come from lead or generator migration, undersensing, a rise in the pacing threshold or a depleted battery. A recording made at the time of a shock helps show which it was, so save the strip, mark the time and note what the patient was doing.
| After a shock | What happens next (NHLBI, Merck) |
|---|---|
| One shock, patient feels well afterward | Provider or device clinic check that day or the next |
| Several shocks in a short time, or chest pain, shortness of breath, fainting | Urgent: tell the nurse or provider at once; outside a hospital, the emergency department |
| Patient unresponsive and not breathing normally | Code team, CPR and the AED as trained |
NHLBI tells patients that a high-energy shock lasts under a second and can be strong or painful, and that a partner touching them usually feels nothing or a tingle that is not harmful. CPR is not delayed for fear of an ICD shock. A patient wearing a defibrillator vest is different: it sounds a warning before it shocks, and anyone touching the patient moves away as soon as it warns.
Around the device: the technician's job
- Before recording, ask about a pacemaker, ICD or other implanted stimulator; look for the scar and bulge below a collarbone.
- Use the diagnostic filter settings for the first recording, and label any filter you add.
- Note the device on the tracing or requisition and tell the reader. If you learn of it after the tracing is filed, add a dated addendum; never change the original entry.
- On telemetry, confirm the monitor's pacer detection is on for every paced patient.
- Spikes that do not fit any pacing pattern: think artifact or a nerve stimulator before malfunction, but report anything you cannot explain.
- AED: do not put a pad over the device. Singapore 2026 advises about four finger breadths away, and notes devices implanted under the right collarbone, where the upper pad usually goes.
- Report, do not decide: magnets, MRI, interrogation and reprogramming belong to the nurse, provider and device team.
Related cheat sheets
- Safety, Compliance and Patient CarePatient ID, consent, HIPAA, infection control, electrical safety, vital signs, recognizing and reporting emergencies, CPR and AED, safe lifting, scope and charting for EKG technicians
- Electrode and Lead PlacementLimb and chest landmarks, counting from the angle of Louis, telemetry and Mason-Likar layouts, right-sided, posterior and 15-lead recordings, children, amputations, dextrocardia, and what to write when placement is not standard
- Artifact and Lead ProblemsTremor, 60-cycle interference, wandering baseline, loose electrodes, lead-off and cable faults, artifact that mimics VT, VF or a pause, and limb and chest lead reversals
- Paper, Settings and MeasurementThe grid, paper speed, gain and the calibration pulse, filters, heart rate methods, PR, QRS and QT in adults and children, QTc formulas and U waves
- Analyzing a Rhythm Strip Step by StepA fixed order for every strip, regularity, narrow vs wide fast rhythms, early vs escape beats, describing what you cannot name, and what to report at once
- AV Blocks, Bundle Branch Blocks and Pre-excitationCriteria tables for every degree of AV block, AV dissociation, bundle branch and fascicular blocks, and WPW, with what to measure and when to report
Sources
- Burns E, Buttner R. Pacemaker Rhythms - Normal Patterns. LITFL ECG Library. Link
- Costello L. Pacemaker Essentials: How to Interpret a Pacemaker ECG. CanadiEM. May 24, 2016. Link
- Mitchell LB. Cardiac Pacemakers. Merck Manual Professional Edition, full review February 2025. Link
- Mitchell LB. Implantable Cardioverter-Defibrillators (ICD). Merck Manual Professional Edition, full review February 2025. Link
- Fernando RJ, Crimi E, Puttur Rajkumar K, et al. Perioperative management of cardiac implantable electronic devices: a state of the art narrative review. J Thorac Dis. 2026;18(5):536. doi:10.21037/jtd-2026-1-0434 Link
- van Stipdonk A, Wijers S, Meine M, Vernooy K. ECG Patterns in Cardiac Resynchronization Therapy. J Atr Fibrillation. 2015;7(6):1214. Link
- Ferrick AM, Raj SR, Deneke T, et al. 2023 HRS/EHRA/APHRS/LAHRS Expert Consensus Statement on Practical Management of the Remote Device Clinic. Europace. 2023;25(5):euad123. doi:10.1093/europace/euad123 Link
- Steinberg JS, Varma N, Cygankiewicz I, et al. 2017 ISHNE-HRS expert consensus statement on ambulatory ECG and external cardiac monitoring/telemetry. Ann Noninvasive Electrocardiol. 2017;22(3):e12447. doi:10.1111/anec.12447 Link
- Drew BJ, Harris P, Zègre-Hemsey JK, et al. Insights into the Problem of Alarm Fatigue with Physiologic Monitor Devices: A Comprehensive Observational Study of Consecutive Intensive Care Unit Patients. PLoS One. 2014;9(10):e110274. doi:10.1371/journal.pone.0110274 Link
- Mondal S, Raja DP, Muslim NA, Prabhu MA. Electrocardiographic artifacts in clinical practice: a logical approach to recognition and prevention. World J Cardiol. 2026;18(3):116299. doi:10.4330/wjc.v18.i3.116299 Link
- Campbell B, Richley D, Ross C, Eggett CJ. Clinical Guidelines by Consensus: Recording a standard 12-lead electrocardiogram (CS3 version 4.1). Society for Cardiological Science and Technology (SCST); September 2024. Link
- Luo S, Johnston P, Hong W. Performance Study of Digital Pacer Spike Detection as Sampling Rate Changes. Computers in Cardiology. 2008;35:349-352. Link
- Okubo K, Nakashima E, Takagi K, Kawaguchi N, Tanaka Y, Takahashi A. Intrinsic antitachycardia pacing for patients with multiple ventricular tachycardias. HeartRhythm Case Rep. 2022;8(10):711-714. Link
- National Heart, Lung, and Blood Institute. Defibrillators: Living With an ICD or WCD. Link
- Lim SH, Leong BSH, Loke JH, Ng KC. Singapore Basic Cardiac Life Support and Automated External Defibrillation Guidelines 2026. Singapore Med J. 2026;67(7):406-413. doi:10.4103/singaporemedj.SMJ-2026-319 Link
- National Healthcareer Association. NHA Certified EKG Technician (CET) Test Plan, based on the 2026 job analysis. Link
- Cardiovascular Credentialing International. Certified Rhythm Analysis Technician (CRAT) Examination Overview. June 2022. Link