Pacemakers and Implanted Devices on the ECG

The pacemaker code, atrial, ventricular, dual-chamber and biventricular pacing, capture and sensing problems, magnets, ICD shocks, and filters that hide or fake spikes

Reviewed by Alex G, BSN, RN, CEN, CCRN, CMC, NREMT-P · Checked against current sources November 2026

Practice 18 Pacemakers and Implanted Devices on the ECG questions free

Page 1 of the Pacemakers and Implanted Devices on the ECG PDF, a printed page in columns with sections on The pacemaker code (NBG), What paced beats look like, Naming the problem, Magnets and ICDs, Filters, monitors and look-alikes and Exam pearlsEnlarge

One-page PDFLetter landscape, ready to print or keep with your notes

View PDFDownload

Pacemakers and Implanted Devices on the ECG: one-page PDF preview

Page 1 of the Pacemakers and Implanted Devices on the ECG PDF, a printed page in columns with sections on The pacemaker code (NBG), What paced beats look like, Naming the problem, Magnets and ICDs, Filters, monitors and look-alikes and Exam pearls

View PDFDownload

The pacemaker code (NBG)

LetterMeansChoices
1stChamber pacedA atrium, V ventricle, D both
2ndChamber sensedA, V, D, O
3rdResponse to a sensed beatI inhibit, T trigger, D both, O none
4thRate responseR raises the rate with activity, O none
5thMultisite pacingFor example both ventricles; O none

VVIR: paces and senses the ventricle, holds back when it senses a beat, speeds up with activity. DDD is the most common mode. AOO, VOO and DOO pace at a fixed rate and sense nothing (asynchronous).

What paced beats look like

PacingOn the strip
SpikeNarrow vertical line, about 2 ms; tiny with bipolar leads, so check every lead
AtrialSpike, then a P wave; the patient's own QRS follows
Ventricular (RV lead)Spike, then a wide QRS shaped like LBBB, ST-T opposite the QRS
Dual-chamberSpikes before the P only, the QRS only, or both, depending on the patient's own beats
Biventricular (CRT)Paced QRS often upright in V1; two ventricular spikes may show as one
FusionSpike on a QRS that is a hybrid of paced and native; normal function
No spikesOwn rhythm is fast enough: a demand pacemaker waits and senses
Lead II · 25 mm/s
Dual-chamber pacing: atrial spike and P wave, a set delay, then a ventricular spike and wide QRS. Ventricular pacing alone, failure to capture and failure to sense are drawn in the Rhythm Gallery.

Naming the problem

ProblemOn the stripCommon causes
Failure to captureSpike with no P wave or QRS after itLead moved or broken, battery, threshold rise; electrolytes, drugs, MI
Undersensing (failure to sense)Spikes ignore the patient's own beats; can land on a T waveLead problem, small native signals
OversensingPause with no spike when one is due; often with muscle noiseT waves, muscle activity or interference read as beats
Failure to paceNo spikes; rate falls below the programmed rateOversensing, battery or generator, poor lead contact
Paced rate too fastPaced rate faster than expected, up to the device's upper rateTracking a fast atrial rhythm, PMT, activity sensor

Magnets and ICDs

DeviceMagnet on (stops when it is lifted)
Most pacemakersFixed-rate pacing at the maker's magnet rate; spikes march through the patient's own beats
Transvenous ICDShocks and antitachycardia pacing suspended; no fixed-rate pacing
Micra leadless pacemakerNo magnet response

Filters, monitors and look-alikes

Exam pearls

Practice Pacemakers and Implanted Devices on the ECG questions free in your browser18 exam-style questions with rationales. No account or download needed.

In depth

The one-page summary above is what the PDF prints. These sections go further.

Spikes: why they are small, missing or fake

A pacing stimulus is very brief, so on paper it is a thin vertical line, usually about 2 ms wide (LITFL). Its height depends on where the lead sits and what kind of lead it is: bipolar leads make much smaller spikes than unipolar leads, and epicardial leads make smaller spikes than leads inside the heart. A spike that is obvious in one lead may be invisible in the next, which is why CanadiEM advises checking every lead before deciding a beat is not paced.

Machines have to work to see them. In one study of electrocardiographs, pacer spike detection from the ordinary 500-samples-per-second data stream found 77.5% of spikes, while a detector sampling at 32,000 per second found 99.5% (Luo 2008). The same study notes that a slow data stream can show only one spike when two ventricular leads fire close together, as in biventricular pacing. Stored ECGs are usually compressed, and SCST warns that compression changes short, sharp signals such as pacemaker spikes more than it changes the ST segment, so an archived copy may not match the original.

Many bedside monitors use a separate pacer detection mode with a higher sampling rate. In Drew's ICU study, when it detected a spike the monitor painted its own marker on the display, so the spike you see may be the monitor's mark, not the real signal (Drew 2014). In that study only one third of patients with pacemakers had the mode switched on, and intermittent ventricular pacing without it produced false 'accelerated ventricular rhythm' alarms. The technician confirms the pacemaker in the chart and asks the nurse to turn the setting on, or turns it on where unit policy allows.

Filter or settingEffect on spikesWhat to do
Diagnostic band, 0.05 or 0.67 to 150 Hz (SCST)Keeps short, sharp signalsUse it for the first recording
40 Hz muscle filterDistorts the ECG; spikes can be blunted or lostOnly after relaxing the patient fails; label the tracing
Any filtering or processingSpikes can be exaggerated or distorted; tall or broad spikes can pass for fibrillation or noise (Mondal 2026)Compare with the patient's other tracings; note the device
Analog-type monitorsDrop expected spikes more often than diagnostic ECGs (Mondal 2026)Confirm on a 12-lead or paper strip before calling a malfunction

Things that make spike-like marks: static electricity, for example from clothing (Luo's test set included spikes from clothing static and a static gun); fractured wires and loose connectors; and other implanted stimulators. Mondal describes neurostimulators (deep brain, sacral nerve and gastric) sending periodic pulses, and deep brain stimulators causing high-frequency artifact, usually 60 to 200 Hz. A stimulator's spikes run at their own fixed rate with no link to the P waves or QRS complexes, and re-prepping the skin does not remove them; ask about implanted devices and tell the nurse or provider. Mondal's rule of thumb: spikes that occur at predictable intervals in relation to atrial or ventricular events are almost always real pacing. Pseudopacing artifacts that no pacing algorithm can explain also occur, and artifact can simulate pacemaker malfunction (ISHNE-HRS), so read every suspected malfunction in context before it is reported as one.

Modes and how a dual-chamber pacemaker behaves

The code follows the NASPE/BPEG (NBG) system, revised in 2002; LITFL notes that many textbooks still print the 1987 version, in which the fourth letter also covered programmability and the fifth covered antitachycardia functions. Positions I to III are the same in both. Merck's table gives position V as O, A, V or D, where multisite pacing means both atria, both ventricles or more than one lead in one chamber; a code can have three to five letters.

ModeWhat it doesWhat the strip shows
AAIPaces and senses the atrium; holds back for the patient's own P waveAtrial spikes only when the sinus rate falls below the set rate
VVI(R)Paces and senses the ventricle; one of the two most used modesVentricular spikes only when the patient's own QRS is late
DDD(R)Senses and paces both chambers; the most common modeNo spikes, atrial spike only, ventricular spike only, or both
AOO, VOO, DOOPaces at a fixed rate and senses nothing; what most pacemakers do under a magnetSpikes march on regardless of the patient's own beats

A DDD pacemaker watches the atrium first. If no P wave comes by the time the lower rate is due, it paces the atrium; then it waits a set delay for the patient's own QRS and paces the ventricle only if none comes (Fernando 2026). That is why LITFL says dual-chamber spikes may precede only the P wave, only the QRS, or both. A P wave with a ventricular spike after it means the device is following the patient's own atrial rate; two spikes per beat means both chambers are paced; no spikes means the patient's own rhythm is in charge.

Atrial pacing can unmask a slow AV node. LITFL shows atrially paced patients with first-degree block or Wenckebach conduction that their slower native rhythm did not show, and calls it clinically unimportant if cardiac output holds. Each atrial spike followed by a P wave proves atrial capture; the dropped beats are still reported to the nurse.

Rate response (R) uses a sensor to speed up pacing with activity. Merck notes that such devices can also respond to vibration and muscle activity, so a bumpy car ride can raise the paced rate; a diary entry beside the faster pacing helps the device clinic. Mode switching protects dual-chamber patients from fast atrial rhythms: in atrial fibrillation, for example, a DDDR device can switch itself to VVIR and pace the ventricle at a programmed rate instead of following the atria. The device is working, but new AF is still a rhythm change to report.

Reading a paced strip step by step

  1. Read the strip as usual, then add four checks (CanadiEM): rate, capture, pacing, sensing.
  2. Rate: measure spike to spike. At 25 mm/s, 1500 divided by the small boxes between spikes is the paced rate (25 small boxes = 60/min). Compare it with the programmed rate in the chart or on the device card.
  3. Capture: is every spike followed by its P wave (atrial) or QRS (ventricular)? A spike that lands just after the patient's own beat finds the heart refractory, so no complex follows; that alone does not prove failure to capture. Look for spikes that fall where the heart could respond.
  4. Pacing: are spikes there whenever the patient's own rate drops below the programmed rate? A pause longer than the pacing interval with no spike is failure to pace, from oversensing or a hardware problem (CanadiEM).
  5. Sensing: do spikes appear when the patient's own beats are already on time? That is undersensing.
  6. Record the device, the findings and the time, and report as below. Leave interrogation, reprogramming and treatment to the nurse, provider and device team.

Fusion and pseudofusion are normal behaviour, not malfunction. When the patient's own rate is close to the programmed rate, the device's timer can run out just as a native beat begins. If both wavefronts share the ventricle, the QRS is a hybrid of the paced and native shapes, often narrower than a fully paced beat (fusion, LITFL and CanadiEM). If the native beat already owns the ventricle, the spike sits on a QRS that it did not change (pseudofusion). Check the QRS shape and timing before calling either one a problem.

FindingHow fast to report
Failure to capture or to pace, pauses, or any finding with dizziness, weakness or faintingNow: stay with the patient and tell the nurse
Undersensing, especially a spike on a T wavePromptly, even without symptoms
Paced rate faster or slower than programmed; new AF in a paced patientPromptly, with the strip
Normal paced rhythm, fusion, or no spikes with a normal native rhythmDocument the device; routine

Malfunctions in more depth

Capture means the stimulus is strong enough to depolarize the heart. CanadiEM groups the causes of failure to capture as hardware (a broken lead or one that has lost contact, battery problems), exit block (a rise in the stimulation threshold where the lead meets the heart), and patient factors (electrolyte problems, drugs, myocardial infarction). Each uncaptured spike is a lost beat, so the true heart rate falls below the programmed rate, which explains weakness or lightheadedness.

Sensing is set by a sensitivity number: a lower number makes the device see smaller signals. Undersensing means the device misses the patient's own beats and paces when it should not, so spikes appear anywhere, from a lead fracture, poor contact, an insulation problem or low-voltage native signals. A spike near the peak of the T wave can cause an R-on-T event, which CanadiEM links to VT or torsades.

Oversensing means the device mistakes something else for a heartbeat and holds back its output: T waves, signals from the wrong chamber, muscle potentials (myopotentials), or electromagnetic interference. The strip shows pauses with no spikes, for example during arm work with muscle noise on the baseline, which separates it from failure to capture (spikes with nothing after them). ISHNE-HRS lists myopotential inhibition, cross-talk and pacemaker-mediated tachycardia as the main abnormalities seen on Holter recordings in pacemaker patients, though they are uncommon, and notes that in one report pacemaker dysfunction explained syncope in only 4.9% of pacemaker patients.

ProblemMechanism (Merck, CanadiEM)On the strip
Cross-talkThe ventricular channel senses the atrial pacing pulse and withholds ventricular pacingAtrial spike and P wave, then no ventricular spike and a pause
Pacemaker-mediated tachycardia (PMT)A PVC or paced beat conducts back to the atria; the atrial lead senses that P wave and the device paces the ventricle again, over and overSudden ventricular-paced tachycardia at or near the upper rate, often after a PVC, with a retrograde P after each QRS
Sensor-driven tachycardiaThe rate sensor responds to vibration or muscle activityPaced rate rises without exertion; match it to the diary
Pacemaker syndromeVentricular pacing that loses the normal timing between atria and ventriclesPaced rhythm that looks normal; the patient has vague dizziness, neck or breathing symptoms. Report them.

PMT needs an atrial lead to sense the retrograde P wave, so it occurs only with dual-chamber devices; the device team can stop and prevent it, so the technician reports it promptly. A patient whose own heart produces little or no escape rhythm is called pacing dependent; pauses from oversensing or cross-talk matter most in these patients.

Biventricular pacing (CRT)

Cardiac resynchronization therapy adds a lead that paces the left ventricle, usually from a coronary vein, so both ventricles are paced together. Pacing from the right ventricular apex alone gives a wide QRS that is negative in V1 (recorded at the 4th intercostal space) with, usually, a left superior axis. During biventricular pacing the QRS most often has a dominant R wave in V1 and V2, and the frontal axis most often points to the right superior quadrant, though other axes do not by themselves mean a problem (van Stipdonk 2015).

A negative QRS in V1 during CRT warrants evaluation. Causes include a V1 electrode placed in the wrong space, slow conduction away from the left ventricular lead (which leaves a picture dominated by right ventricular pacing) and a dislodged left ventricular lead. The technician's first job is to confirm that V1 sits in the 4th intercostal space at the right sternal edge, then to report a change in the paced QRS shape compared with earlier tracings.

CRT helps only when it paces nearly every beat. ISHNE-HRS says the benefit depends mostly on effective biventricular pacing, and that programming should aim for the maximum, above 98%. The 2023 HRS/EHRA/APHRS/LAHRS remote device clinic consensus notes that studies have used cut-offs from above 80% to above 98% and names above 95% as the target. PVCs, AF and fusion or pseudofusion beats all lower the true share. In one small study of CRT patients in AF whose devices counted over 90% biventricular pacing, a 12-lead Holter found fusion or pseudofusion beats in 53% of the patients, so a high device count can overstate real pacing. A high PVC or AF burden on a Holter is a finding the device clinic needs.

Magnet response in more depth

Most pacemakers switch to asynchronous pacing (AOO, VOO or DOO, depending on the leads) while a magnet is over them, at a rate set by the maker. Sensing is off, so spikes compete with the patient's own beats, and spikes that land just after a native beat find the ventricle refractory; that is expected while the magnet is on. LITFL notes that asynchronous ventricular pacing carries a risk of pacemaker-induced VT, so a spike on a T wave is still reported. The magnet rate is lower once the battery reaches its elective replacement point (ERI), by about 20% (Fernando 2026).

Exceptions exist, and magnet mode can be programmed off. Biotronik devices respond for 8 hours unless the magnet is lifted and replaced, and in their default auto mode give 10 asynchronous beats and then return to normal settings.
Maker (Fernando 2026, general responses)Normal magnet rateAt replacement (ERI)
Abbott (includes St. Jude)100/min85/min
Biotronik90/min80/min
Boston Scientific100/min85/min
Medtronic85/min65/min
Microport96/min80/min

ICDs: pacing, antitachycardia pacing and shocks

A transvenous ICD has leads in the right ventricle and sometimes the right atrium (a CRT-D adds a left ventricular lead). It cardioverts or defibrillates VT and VF, and tiered-therapy ICDs also pace for slow rhythms and use antitachycardia pacing (ATP) on tachycardias that respond to it; they store intracardiac electrograms of each event (Merck). Its bradycardia pacing looks the same on the ECG as a pacemaker's.

ATP delivers a burst of stimuli slightly faster than the tachycardia; in Okubo's cases the burst interval was 88% of the VT cycle length. On a monitor or Holter it appears as a short run of rapid paced beats interrupting the VT, followed by the return of the patient's rhythm if it worked. It is painless, but it fails more often in fast VT (success as low as 50% when the cycle length is under 300 ms), and it can accelerate VT in up to 10% of therapies, after which the device may shock.

Merck lists the causes of inappropriate shocks or pacing: sinus rhythm, supraventricular tachycardia, atrial fibrillation or flutter, and nonphysiologic signals such as those from a fractured lead. Missed therapy can come from lead or generator migration, undersensing, a rise in the pacing threshold or a depleted battery. A recording made at the time of a shock helps show which it was, so save the strip, mark the time and note what the patient was doing.

After a shockWhat happens next (NHLBI, Merck)
One shock, patient feels well afterwardProvider or device clinic check that day or the next
Several shocks in a short time, or chest pain, shortness of breath, faintingUrgent: tell the nurse or provider at once; outside a hospital, the emergency department
Patient unresponsive and not breathing normallyCode team, CPR and the AED as trained

NHLBI tells patients that a high-energy shock lasts under a second and can be strong or painful, and that a partner touching them usually feels nothing or a tingle that is not harmful. CPR is not delayed for fear of an ICD shock. A patient wearing a defibrillator vest is different: it sounds a warning before it shocks, and anyone touching the patient moves away as soon as it warns.

Around the device: the technician's job

Practice free in your browser650+ exam-style questions with rationales. Free today, no account, no ads. Pro is coming in January 2027. Also free on Google Play.

Related cheat sheets

Sources

  1. Burns E, Buttner R. Pacemaker Rhythms - Normal Patterns. LITFL ECG Library. Link
  2. Costello L. Pacemaker Essentials: How to Interpret a Pacemaker ECG. CanadiEM. May 24, 2016. Link
  3. Mitchell LB. Cardiac Pacemakers. Merck Manual Professional Edition, full review February 2025. Link
  4. Mitchell LB. Implantable Cardioverter-Defibrillators (ICD). Merck Manual Professional Edition, full review February 2025. Link
  5. Fernando RJ, Crimi E, Puttur Rajkumar K, et al. Perioperative management of cardiac implantable electronic devices: a state of the art narrative review. J Thorac Dis. 2026;18(5):536. doi:10.21037/jtd-2026-1-0434 Link
  6. van Stipdonk A, Wijers S, Meine M, Vernooy K. ECG Patterns in Cardiac Resynchronization Therapy. J Atr Fibrillation. 2015;7(6):1214. Link
  7. Ferrick AM, Raj SR, Deneke T, et al. 2023 HRS/EHRA/APHRS/LAHRS Expert Consensus Statement on Practical Management of the Remote Device Clinic. Europace. 2023;25(5):euad123. doi:10.1093/europace/euad123 Link
  8. Steinberg JS, Varma N, Cygankiewicz I, et al. 2017 ISHNE-HRS expert consensus statement on ambulatory ECG and external cardiac monitoring/telemetry. Ann Noninvasive Electrocardiol. 2017;22(3):e12447. doi:10.1111/anec.12447 Link
  9. Drew BJ, Harris P, Zègre-Hemsey JK, et al. Insights into the Problem of Alarm Fatigue with Physiologic Monitor Devices: A Comprehensive Observational Study of Consecutive Intensive Care Unit Patients. PLoS One. 2014;9(10):e110274. doi:10.1371/journal.pone.0110274 Link
  10. Mondal S, Raja DP, Muslim NA, Prabhu MA. Electrocardiographic artifacts in clinical practice: a logical approach to recognition and prevention. World J Cardiol. 2026;18(3):116299. doi:10.4330/wjc.v18.i3.116299 Link
  11. Campbell B, Richley D, Ross C, Eggett CJ. Clinical Guidelines by Consensus: Recording a standard 12-lead electrocardiogram (CS3 version 4.1). Society for Cardiological Science and Technology (SCST); September 2024. Link
  12. Luo S, Johnston P, Hong W. Performance Study of Digital Pacer Spike Detection as Sampling Rate Changes. Computers in Cardiology. 2008;35:349-352. Link
  13. Okubo K, Nakashima E, Takagi K, Kawaguchi N, Tanaka Y, Takahashi A. Intrinsic antitachycardia pacing for patients with multiple ventricular tachycardias. HeartRhythm Case Rep. 2022;8(10):711-714. Link
  14. National Heart, Lung, and Blood Institute. Defibrillators: Living With an ICD or WCD. Link
  15. Lim SH, Leong BSH, Loke JH, Ng KC. Singapore Basic Cardiac Life Support and Automated External Defibrillation Guidelines 2026. Singapore Med J. 2026;67(7):406-413. doi:10.4103/singaporemedj.SMJ-2026-319 Link
  16. National Healthcareer Association. NHA Certified EKG Technician (CET) Test Plan, based on the 2026 job analysis. Link
  17. Cardiovascular Credentialing International. Certified Rhythm Analysis Technician (CRAT) Examination Overview. June 2022. Link